Steam Sterilizer Operation Manual Please read the manual carefully, operate it according to the manual strictly, and avoid any hazard! This manual applies to SUN8-Ⅲ、SUN12-Ⅲ、SUN16-Ⅲ、SUN18-Ⅲ、 SUN23-Ⅲ (M)Steam Sterilizer. File No:MTQM/C1-7.5.1-05-02 REV1607ER...
Steam Sterilizer 1.General Introduction This Steam Sterilizer is a precision instrument specializing in the operation occasion in need of simple, fast and efficient sterilization by the doctor or his/her assistants, and the doctor shall take the responsibility for such work. This Steam Sterilizer is specially designed for the occasions in use of frequent sterilizations.
person and equipment. The chloride ion is an important factor of stainless steel corrosion da mage. If use this sterilizer sterilization items containing chlorine ion, must rinse sterilization indoor wall every day, in order to avoid deposition on i nternal stainless steel corrosion of chlorine ion, delayed the service life of the product.
4.Diagram Specification Symbol Meaning CAUTION Warning: dangerous voltage CE MARKING, with notified body number 0197 WEEE MARKING MANUFACTURER DATE OF MANUFACTURE AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY SERIAL NUMBER CONSULT INSTRUCTIONS FOR USE Earthing 5.Installation 5.1 Remove the packing Dismantle thin film, open the packing,get the equipment from the packing Hauling precautions: ·forbid lift up door ·forbid lift up legs of machine...
Page 6
5.2 The air-cooling window attached to the exterior surface of the Steam Sterilizer shall be non-clog by dust or articles, and the Steam Sterilizer shall working in the proper air flowing environment,and no dangerous flammable and explosive gas. Equipment normal working conditions: environment temperature:+5~+40℃...
The equipment must be reliable grounding. Must use distilled water Attention! Through poor conditions of the electrical MAINS, shortly voltage drops can appear when starting the EQUIPMENT. This can influence other equipment (eg. Blinking of a lamp). If the MAINS-IMPEDANCE Zmax 0.114OHM, such disturbances are not expected.
7. Display and keypad description 7.1 Control Panel 1)Temperature display window Displaying the temperature inside the chamber. Unit: ºC 2)Status/ Error code number display window Applied for displaying kinds of status of the autoclave during its working time. See more details in “Appendix working status display form”.When alarm occurs to the autoclave, the digital pipe of the display will show the alarm code number, the user can find out the malfunction part.(Once the malfunction occurring to the autoclave, please contact the service center or distributor immediately).
Page 9
use UP and Down to change the data, then press START to quit and save the setting. Press SET to enter the advanced settings status, press SET again to set the custom setting of the working cycle, like the Temperature, Sterilizing time..., use UP and Down to change the data you want to set, then press START to quit and save the setting.
Page 10
When the machine displaying the used water full, after sterilizing, please drain used water out in time. Please clean the brass filter inside the chamber, or it would effect the pressure releasing during the sterilization. Please clean the filter in the same time with you draining the used water.
7.3.4 Exhaust drying process: When the sterilization time, reduced to zero, into the exhaust stroke dry, open the drain valve to drain the pressure reduced to +0.0then the vacuum pump to start. When the drying time only 3 minutes ,open the intake valve to vacuum drying. 7.3.5 Pressure equilibrium process: When the drying time is reduced to zero, turn off the exhaust, dry intake valve opens, so that the pressure to achieve and external balance, open the automatic...
Page 12
Must wait after the pressure release to “0 Kpa” , then open the door! Even the pressure is 0 Kpa, there is still hot steam inside the chamber, Operator need to take care it and prevent scald! Error code and resolving measures Code Alarm Tone Malfunction content...
9. Maintenance and Service Shut off the power before service or components’ replacement, and the service or replacement work shall be done by the technician. Working environment minimum brightness to 215 + 15 lx. 9.1 Clean the top water tank by disinfectant or alcohol once a week. 9.2 Disinfect and clean the interior surface of the Steam Sterilizer chamber monthly.
4)Note: The inner circle of the ring may be raised up when insert the ring into the groove. Please press it to the groove by using the screwdriver carefully. Shut off the power and cooling the Steam Sterilizer sufficiently before inserting the ring for avoiding scald.
collecting tank, 4 is the water exiting mouth of the water storing tank. Revolving the water exit button by anticlockwise direction so as to release the water. 10.3 The autoclave shall be transferred and stored with the below conditions: Ambiance temperature: -40ºC -+55ºC Relative humidity: <...
Page 16
cavity body is 125 ℃.When choose 134 ℃ sterilization procedure, the highest temperature in cavity body is 138 ℃. 19.This product is not suitable for sealing, not resistant to high temperature products sterilization. 20. This product is not suitable for airtight liquid class sterilization 21.The product is prohibited to sterilized the liquid which were sealed into glass bottle or glass vessel, Because when operating, the change of temperature and pressure may cause the liquid bottle burst, so that endanger the safety of the...
APPENDIX NO.1: The working status display table: Digital display Working status name Working status specification Prepare/preheating status In this status, prepare to start working. Heating status In this status, the pressure and temperature increasing ± pressure data Displaying the ±pressure In this status, the chamber have working pressure Displaying the...
Page 19
NO.4:Vacuum test "B&D" test and Helix text methods Press UP / DOWN key to choose the test system, close the door, then press "START" button to start. Ningbo Mingtai Medical Instrument Co., Ltd Yunzhen Road, Industrial Zone, Yunlong Town, Yinzhou, Ningbo City, China Tel: +86-0574-88320096;Fax:+86-0574-88320098...