Steam Sterilizer
Operation Manual
Please read the manual carefully, operate it according to the manual
strictly, and avoid any hazard!
This manual applies to SUN18-Ⅲ、SUN23-Ⅲ (S)Steam Sterilizer.
File No:MTQM/C1-7.5.1-05-02/Ⅲ-S(E )10
REV1607ER

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SUN18-III and is the answer not in the manual?

Questions and answers

Summary of Contents for MingTai SUN18-III

  • Page 1 Steam Sterilizer Operation Manual Please read the manual carefully, operate it according to the manual strictly, and avoid any hazard! This manual applies to SUN18-Ⅲ、SUN23-Ⅲ (S)Steam Sterilizer. File No:MTQM/C1-7.5.1-05-02/Ⅲ-S(E )10 REV1607ER...
  • Page 2: Table Of Contents

    Content 1、General Introduction ……………………………………….1 2、Application Scope ………………………………………….1 3、Tech. Parameter…………………………………………….1 4、Diagram Specification ………………………………………2 5、Installation …………………………………………………..3 6、Packing accessory…………………………………………5 7、Display and keypad description ……………………………5 8、Emergency Condition……………………………………….9 9、Maintenance and Service……………………………………9 10、 Transportation and Storage…………………………………11 11、 Attention……………………………………………………12 Appendix: NO.1: Circuit diagram of sterilizer NO.2: Steam loop...
  • Page 3: 1、General Introduction

    Steam Sterilizer 1.General Introduction This Steam Sterilizer is a precision instrument specializing in the operation occasion in need of simple, fast and efficient sterilization by the doctor or his/her assistants, and the doctor shall take the responsibility for such work. This Steam Sterilizer is specially designed for the occasions in use of frequent sterilizations.
  • Page 4: 3、Tech. Parameter

    person and equipment. The chloride ion is an important factor of stainless steel corrosion da mage. If use this sterilizer sterilization items containing chlorine ion, must rinse sterilization indoor wall every day, in order to avoid deposition on i nternal stainless steel corrosion of chlorine ion, delayed the service life of the product.
  • Page 5: 5、Installation

    WEEE MARKING MANUFACTURER DATE OF MANUFACTURE AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY SERIAL NUMBER CONSULT INSTRUCTIONS FOR USE Earthing 5.Installation 5.1 Remove the packing Dismantle thin film, open the packing,get the equipment from the packing Hauling precautions: ·forbid lift up door ·forbid lift up legs of machine ·forbid Rollover or upside down ·Hauling sketch map is as follows:...
  • Page 6 proper air flowing environment,and no dangerous flammable and explosive gas. Equipment normal working conditions: environment temperature:+5~+40℃ Relative humidity:≤85% barometric pressure:70kPa~106kPa minimum brightness: 215±15 lx, 5.3 The Steam Sterilizer shall be installed on the horizontal working base. Put machine on the table, refer to specifications, table must be able to bear the weight of the machine.
  • Page 7: 6、Packing Accessory

    Beim Anlaufen (Starten) dieser Ausrüstungen kann ein kurzzeitiger Spannungseinbruch auftreten, insbesondere bei schlechter Netzqualität. Diese Einbrüche können andere Geräte beeinflussen (z.B. Flimmern einer Lampe). Bei einer Netzimpedanz Zmax 0.114 OHM sind solche Störungen nicht zu erwarten. (Bitte kontaktieren Sie Ihre lokales Energieversorgungsunternehmen für weitere Informationen).
  • Page 8: 7、Display And Keypad Description

    7. Display and keypad description 7.1 Display Interface 7.1.1 Starting Interface:Press Enter to start 7.1.2 Menu Interface:interchange the different interfaces with left buttons 7.1.3 Operation Interface: ·Temp T : Current temp inside of the Chamber.
  • Page 9 ·Pressure P: Current pressure insdie of the Chamber. ·Status ( at the position of icon “door closed”): At normal working time,dispaly the standard cycling status. · Mulfuntion Code( Icon “Water Scarce” ): Alarming with different error codes (see the codes index) . 7.1.4 Setting Interface: ·Water detect: Enable the fuction or not.
  • Page 10 7.2 Fill & drain water When you install the Steam Sterilizer for the first time, after turn on the switch, then The machine would displaying lack of the source water. Then you need to use the measuring cup to fill distilled water from the top inlet around 4000ml for the first time, next time do not filling the water too full.
  • Page 11 7.3.2 vacuum process: Chamber wall temperature to 50 ℃, when the steam generator temperature reaches 100 ℃, into the vacuum heating process, open the vacuum selection valve, exhaust valve (normally on) and start the vacuum pump, 300 seconds later Close the vacuum selection valve, exhaust valve (normally on) and the vacuum pump to stop pumping the vacuum.
  • Page 12: 8、Emergency Condition

    Sterilizing 210KPa 210KPa 110KPa 110KPa 210KPa pressure Vacuum times Sterilizing time Drying time 7.4 Emergency exit, Long press the start key 5 s, Machine will out of sterilization procedures, "EC" alarm meanwhile 7.5 After sterilization, use handheld we provide accessories to take devices Must use handheld to take things, prevent burns! 8.
  • Page 13: 9、Maintenance And Service

    keep it clean and no jam long "du" Problem of door switch or door Inspect door switch or if door being opened when working handle were revolved to the right MAX position long "du" Fault of temperature senor on steam Inspecting the temperature sensor generator on steam generator...
  • Page 14 9.4 Clean the sealing ring periodically The operator shall clean the sealing ring periodically for avoiding the influence to the seal caused by the dust/dirt left due to the long period use. Use the smooth cloth with distilled water to wipe the surface of the sealing ring or sealing cap gently.
  • Page 15: 10、 Transportation And Storage

    2) replace with a qualified safety valve (must provide by manufacturers). 3) Test a sterilization process The machine maintenance and replacement of electrical components must cut off power supply first. And only the manufacturer or professional person designated by the manufacturer can operate 10.
  • Page 16 the cooling water collecting tank shall be emptied once the water storing tank emptied. 6. The door handle must be buttoned up before working. 7. Don’t open the door of Steam Sterilizer before seeing the “0” kPa displayed on the pressure indicator. 8.
  • Page 17 APPENDIX NO.1:Circuit diagram...
  • Page 18 NO.2:Steam Loop...
  • Page 19 Ningbo Mingtai Medical Instrument Co., Ltd Yunzhen Road, Industrial Zone, Yunlong Town, Yinzhou, Ningbo City, China Tel: +86-0574-88320096;Fax:+86-0574-88320098 Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany Tel: +49-40-2513175 Fax :+49-40-255726...