Mechanical system
Fig. 6:
•
Carefully remove grid from grid holder. The gaskets are no longer needed.
•
Apply some silicon (20 48 981) to the fitting surface of the new metal grid holder (75 51
703), all the way around, orient the grid correctly, and press it onto the grid holder.
Siemens AG
Medical Solutions
I.I._3
SPR2-230.814.03.02.02
11.05
SIREMOBIL Iso-C 3D
Page 19 of 52
CS SD 24
19