Guidance and manufacturer's declaration - electromagnetic
immunity
The Stability Pro is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure that it
is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-11
IEC 60601 test
level
±8 kV contact
±15 kV air
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 1 kV line(s)
to line(s)
± 2 kV lines(s)
to earth
<5 % U
T
(>95 % dip in
U
) for 0,5
T
cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
<5 % U
T
(>95 % dip in
U
) for 5 s
T
Compliance
Electromagnetic
level
environment - guidance
Compliant
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Compliant
Mains power quality
should be that of a typical
commercial or hospital
environment.
Compliant
Mains power quality
should be that of a typical
commercial or hospital
environment.
Compliant
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the Stability
Pro requires continued
operation during power
mains interruptions, it is
recommended that the
Stability Pro be powered
from an uninterruptible
power supply or a battery.