Calibration procedure

Advertisement

3.5 High Frequency CMRR adjustment
b. Referring to Fig. 11, connect output BNC to the oscilloscope has good grounding.
c. Connect both input lines to a 155 Vp-p, 20 kHz, sine wave signal source.
d. Put SW1 at 1/100.
e. Set VOLTS/DIV switch on oscilloscope's panel to 1 mV and the TIME/DIV switch to 0.1ms.
f. Adjust VC1 slightly to make the amplitude of displayed waveform as small as possible.
Finally the amplitude should be less than 2 mVp-p (after excluding noise).
3.6 Repeat 3.2 and 3.5 adjustment until both adjustments are O.K.
Fig. 11
7