Table A-2 Dc Output Power - Racal Instruments 1264C User Manual

6-slot vxibus mainframe
Table of Contents

Advertisement

Characteristic
Maximum Power Available
To Modules
Useable Power
DC Current Capacity (I
DC Voltage Regulation
Maximum Load Ripple/Noise
Maximum Induced
Ripple/Noise
Dynamic Current I
Protections
Specifications A-2
Table A-2, DC Output Power
Description
500 W, 0
Derate 2.5%/
790W
)
Voltage
MP
+24V
+12 V
+5 V
-2V
-5.2 V
-12 V
-24 V
Voltage
+24 V
+12 V
+5V
-2V
-5.2 V
-12 V
-24 V
Voltage
+24 V
+12 V
+5 V
-2 V
-5.2 V
-12 V
-24 V
Voltage
+24 V
+12 V
+5 V
-2 V
-5.2 V
-12 V
-24 V
Voltage
dm
+24
+12
+5
–2
-5.2
–12
–24
Over voltage protection
Over temperature protection
Over current protection
Short circuit protection
o
o
C to 50
C,
o
o
C above 50
C
I
(Steady-State Current)
MP
4 A
96W
8 A
96W
45 A
225W
15 A
30W
30 A
150W
8 A
96W
4 A
96W
Tolerance, _V
+1.2V, -720 mV
+600 mV, -360mV
+250 mV, -125 mV
-100 mV, +100 mV
-260 mV, +156 mV
-600 mV, +360 mV
-1.2 V, +720 mV
Ripple/Noise
150 mV
pp
50 mV
pp
50 mV
pp
50 mV
pp
50 mV
pp
50 mV
pp
150 mV
pp
Ripple/Noise
150 mV
pp
50 mV
pp
50 mV
pp
50 mV
pp
50 mV
pp
50 mV
pp
150 mV
pp
I
Dynamic Current
dm
3
2
5
3
4
2
3
Racal Instruments © 1998
1264C User Manual

Advertisement

Table of Contents
loading