Baoding Longer Precision Pump LEAD-2 Longer Operating Manual

Peristaltic pump

Advertisement

Quick Links

LEAD - 2 PERISTALTIC PUMP
OPERATING MANUAL
Baoding Longer Precision Pump Co.,Ltd.

Advertisement

Table of Contents
loading

Summary of Contents for Baoding Longer Precision Pump LEAD-2 Longer

  • Page 1 LEAD - 2 PERISTALTIC PUMP OPERATING MANUAL Baoding Longer Precision Pump Co.,Ltd.
  • Page 2 LEAD - 2 OPERATING MANUAL IMPORTANT INFORMATION: Please read operation manual carefully before operation. WARNINGS: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to protect operator and equipment. Please check the tubing frequently and change the tubing in time. If the power line or the plug is worn or damaged please pull out the plug (Hold the plug not the power line when pulling out).
  • Page 3: Table Of Contents

    Table of Contents Introduction Operation Panel Basic Operation Start/Stop Key Direction Key Adjusting Knob Prime Key Running Interface Menu Frame Work Function Setting Calibration System Setting Control Mode Exit Acceptable Pump Heads and Tubing Pump Head Mounting Tubing Loading Remote Control Function Optional Accessories Maintenance Warranty...
  • Page 4: Introduction

    LEAD - 2 OPERATING MANUAL Introduction Lead-2 peristaltic pump is an accurate flow rate pump. Its acceptable pump heads are 15-48, 15-88, 15-24 and 15-44. 2, 4 and 8 channels are available. It offers flow rates of 0.001 300 mL/min. It has flow rate and speed adjusting function.
  • Page 5: Operation Panel

    LEAD - 2 OPERATING MANUAL Operation Panel LCD Screen Prime Key Direction Key Start/Stop Key Adjusting Knob Basic Operation Start/Stop Key Press the Start/Stop Key to start or stop the pump. Direction Key Press the direction key to change the rotation direction of the pump.
  • Page 6: Running Interface

    LEAD - 2 OPERATING MANUAL Running Interface Display running interface after initialization. 123.4mL/min 90.5 rpm ■ ■ ID 0.5 ID 0.5 Flow Rate Display Speed Display Running direction. Current flow rate and speed. Tubing inner diameter. “ID0.5” means the inner diameter of the tubing is 0.5 mm.
  • Page 7: Function Setting

    LEAD - 2 OPERATING MANUAL Function Setting In running interface, press the Adjusting Knob to enter the calibration interface. Turn the Adjusting Knob to select 4 function setting interfaces. Running interface Turn the Press 123.4mL/min Knob the Knob 2. ystem Setting 1.Calibration ■...
  • Page 8 LEAD - 2 OPERATING MANUAL Flow Rate Calibration: 1. For the first time flow rate calibration, “System Setting” must be finished. Please see page 6 for setting method. 2. Press the Start/Stop Key to begin the testing. Press the Start/Stop Key again to stop the testing.
  • Page 9: System Setting

    LEAD - 2 OPERATING MANUAL System Setting The submenus of “System Setting” menu are “a. Tubing ID”, “b. Test time”, “c. Remote Control”, “d. Pump I.D. #”, “e. Footswitch” and “f. Back”. Running interface Press Turn the 123.4mL/min the Knob Knob 1.Calibration 2.
  • Page 10 LEAD - 2 OPERATING MANUAL b. Test time The test time is the required time for one time flow rate calibration. In “System Setting” interface, press and then turn the Adjusting Knob to enter “b. Test time” interface. Press the Adjusting Knob to enter “Set test time”...
  • Page 11 LEAD - 2 OPERATING MANUAL d.Pump I.D. When control computer controls many pumps through RS485 interface, it must identify each pump's I.D.. This pump I.D. should be unique. It's the identification of this pump. In “System Setting” interface, press and then turn the Adjusting Knob to enter “d.
  • Page 12: Control Mode

    LEAD - 2 OPERATING MANUAL Note: The pump identifies the Longer Footswitch automatically. When the pump connects the footswitch, the Start/Stop Key operating panel is invalid no matter the remote control is on or off. f. Back Return back to “System Setting” interface after all the parameters in “System Setting”...
  • Page 13: Exit

    LEAD - 2 OPERATING MANUAL Exit Return back to running interface after all the parameters are set through this interface. In “Running Interface”, press and then turn the Adjusting Knob to enter “4. Exit” interface. Press the Adjusting Knob to return back to “Running Interface”.
  • Page 14: Pump Head Mounting

    LEAD - 2 OPERATING MANUAL Pump Head Mounting The pump head is mounted on drive before the pump leaves the factory. Follow below procedure for changing another pump head. M6 Screw Cartridge Pump Head Mounting Hole for Pump Head Drive 1.
  • Page 15: Tubing Loading

    LEAD - 2 OPERATING MANUAL Tubing Loading Trigger Trigger 1. Press the trigger according to the arrow head directions. Take out the cartridges. 2. Load tubing with stops in the c a r t r i d g e s . M o u n t t h e cartridges on the pump head.
  • Page 16: Remote Control Function

    LEAD - 2 OPERATING MANUAL Remote Control Communication Control - RS485 Remote Control DB15 plug 4 - 20 mA speed control Analog Control 0 - 5 V speed control 0 - 10 V speed control Communication Control The pump can connect to control computer (computer, PLC, SCM) through RS485 serial communication module (shown as below).
  • Page 17 LEAD - 2 OPERATING MANUAL Terminal Definition of Standard External Control Module Analog speed control input, for external speed control in external control modules of 4 to 20 mA, 0 to 5V and 0 to 10 V. External start/stop control input. When connected to COM, the pump runs, when connected to high level (5 to 12 V), the pump stops.
  • Page 18: Optional Accessories

    LEAD - 2 OPERATING MANUAL Optional Accessory Footswitch Connect to the external control interface. Control the start and stop of the pump. Note: Footswitch is optional accessories. It can only control the start and stop of the pump. The speed and direction of the pump are controlled by the knob and switch on the operation penal.
  • Page 19: Technical Specifications

    LEAD - 2 OPERATING MANUAL Technical Specifications Functions Applicable Pump Head: LEAD15-24, LEAD15-44, LEAD15-48, LEAD15-88 Flow Rate Adjustment: Realize accurate flow rate adjustment function Flow Rate Calibration: Calibrate the setting flow rate to get accurate flow rate Speed Control: Realize accurate speed control through rotary coded switch Display: 128 32 graphic LCD displays current running status ×...
  • Page 20 Baoding Longer Precision Pump Co.,Ltd ADD: Building A,Chuangye Center Baoding National High - Tech Industrial Development Zone Baoding,Hebei,China 071051 TEL: 86 - 312 - 3110087 3138553 FAX: 86 - 312 - 3168553 E - mail: longer@longerpump.com Www.longerpump.com...