Digital Setup - Global Water GL500-2-1 Manual

Table of Contents

Advertisement

Global Water
800-876-1172 • globalw.com
Click on the Low EU field and enter the
low engineering unit that will be used
to calibrate the logger, not necessarily
the lowest reading the sensor can output.
In this example, the value is zero feet, which
is the lowest calibration point and water level.
Click Next.
Place the sensor in the condition that will
cause it to output the level corresponding
to the low EU entered in the previous menu.
Wait for the data reading to stabilize.
Click Next to complete the calibration.
XV. Digital Setup
The Digital Setup screen is used for setting up the digital pulse channel.
- 25 -

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gl500u-2-1Gl500s-2-1Gl500-7-2Fr0000k

Table of Contents