Positioning The Unit Correctly; Measuring; Standard Mode - Tanita ViScan AB-140 Instruction Manual

Abdominal fat analyser
Table of Contents

Advertisement

Before You Start

Positioning the Unit Correctly

Use on a flat, stable surface.
• Do not place in direct sunlight, near heaters or the air stream from air-conditioners.
• Do not place in locations of extreme temperature fluctuations.
• Do not place in highly humid or damp places.
For Your Reference
Store the impedance meter in the main unit when not in use.
1. Slide the impedance
meter down from the
top.
Do not expose the impedance meter to strong impacts when it is stored in the main unit as
it may fall off.
12
cont'd
2. Hang the switch
hanger on the
main unit hook
(ensure that the
Power LED on the
impedance meter
is off).
Check that the
Power LED is off.
Please...

Measuring

standard mode

1
Have the subject lie down facing up on a thin mat or
examination table.
2
Expose the navel area.
3
Position the main unit over the abdominal area of the
supine subject.
*If the patient is clothed, completely expose the skin on abdomen and sides for
an area of 10cm centred on the navel.
*The display of the main unit may be on either the left or right side of the subject.
Expose the skin in this area.
Suitably firm surface
(thin mat, examination table, etc.).
Please...
Ensure he subject's hands are placed on their chest as
shown in the diagram above.
As the distance sensor measures the breadth of the
abdomen, ensure that no clothes or elbows impinge on
the 10cm area centred on the navel.
Do not allow the distance sensors to come into contact with the abdomen. Refer to page
21, Waist Circumferences Over 130cm, when contact is unavoidable due to a large waist,
or when the main unit cannot accommodate the size of the abdomen.
Pillows may cause measurement errors. Avoid using them as much as possible.
The setting up phase is now complete.