UNI-T UT323 Operating Manual
Hide thumbs Also See for UT323:

Advertisement

Quick Links

Overview
The Uni-Trend Model UT321, UT322, UT323 and UT325 Thermometers
('the thermometer) are microprocessor-based, digital thermometers designed
to use external J-, K-, T-, E-, R-, S- and N- type thermocouples (temperature
probes) as temperature sensors.
UT321: single input(T1), suitable for K-, J-, T- and E- type thermocouples
UT322: dual input(T1,T2), suitable for K-, J-, T- and E- type thermocouples
UT323: single input(T1), suitable for K-, J-, T-, E-, R-, S- and N- type thermocouples.
Equipped with over limit alarm, over limit signal output and user self-debug
features.
UT325: dual input(T1,T2), suitable for K-, J-, T-, E-, R-, S- and N- type
thermocouples.
Equipped with over limit alarm, over limit signal output and user self-debug features.
Safety Information
Use the thermometer only as specified in this manual. Otherwise, the protection
provided by the meter may be impaired.
Refer to safety information in Table 1 and the international symbols in Table 2.
Table 1. Safety Information
Warning
A warning identified conditions and actions that pose hazards to the user. To avoid
electrical shock or personal injury, follow these guidelines:
Before using the thermometer inspect the case. Do not use the thermometer if it
appears damaged. Look for cracks or missing plastic. Pay particular attention to
the insulation around the connectors.
Disconnect the thermocouple(s) from the thermometer before opening the case.
Replace the batteries as soon as the battery indicator (
possibility of false readings can lead to personal injury.
Do not use the thermometer if it operates abnormally. Protection may be
impaired. When in doubt, have the thermometer serviced.
Do not operate the thermometer around explosive gas, vapor, or dust.
Do not apply more than the rated voltage, as marked on the thermometer (30V),
between the thermocouple(s), or between any thermocouple and earth ground.
When potential differences are anticipated between the thermocouples, use
electrically insulated thermocouples.
When servicing the thermometer, use only specified replacement parts.
Do not use the thermometer with any part of the case or cover removed.
Caution
A caution identifies conditions and actions that may damage the meter or the
equipment under test.
Use the proper thermocouples, function, and the range of your thermometer.
When carrying two thermocouples measurement, make sure there are no
potential differences between two thermocouples.
Do not attempt to recharge the battery.
Match the + and – polarities of the battery with the battery case.
Table 2. International Symbols
Refer to the manual for information about this feature
Complies with European Union directives
Battery
Get Acquainted with Your Thermometer
Unless otherwise specified, all the descriptions apply to all Models of UT321,
UT322, UT323 and UT325.
To become familiar with the thermometer, study the following:
Figure 1 and Table 3 for components;
Figure 2 and Table 4 for LCD display;
Figure 3-a and 3-b and Table 5 for button functions;
Figure 1. Components
Table 3. Components
1
Display
2
Buttons
3
Thermocouple T1 input
4
Thermocouple T2 input (UT323 and UT325 only)
5
USB Port
6
SIGN Port - Over Limit Signal Output (UT323 and UT325 only)
7
NTC
8
Battery Door
9
9V battery (6F22 )
LCD Display
Table 4. LCD Display
1
Setup is in progress when the icon blinks
2
Display readings of maximum, minimum and average
3
Data Transferring is in progress
4
Logged readings are displayed when the icon blinks
5
Low battery indicator. Replace the battery
6
The thermocouple type
7
The temperature units
8
Secondary Display 2
9
Under Calibration mode when the icon blinks. The displayed reading is fixed.
10
Secondary Display 1
The thermocouple measurement includes an offset. See "Changing Setup
11
Options"
12
Readings are being logged when the icon blinks
13
Primary Display
14
) appears. The
Data hold function is active.
The symbol of display backlight
Figure 3-a UT321 and UT323 Keypad
Table 5. Buttons
Press to turn the thermometer on or off
Press to turn the display backlight on and off.
HOLD
Press to freeze or unfreeze the displayed readings
ºC ºF
Press to switch between Celsius (ºC), Fahrenheit(ºF), and Kelvin (K)
Press to step through the maximum, minimum, and average
MAX MIN
readings. Press and hold to exit this function.
Press to step through K-, J-, T-, E- (R-, S-, N-) type thermocouple.
Press to select and show T1, T2 and T1-T2 (differential temperature
measurement) in the primary or secondary display 1
Press to enter USB mode and the USB icon blinks. Press again
SEND
to exit USB mode.
Press to start or stop logging.
STORE
(See "Using Memory - Starting and Stopping Logging".)
Press to show logged readings
RECALL
Press again to stop.
Figure 2. Display LCD Display
Figure 3-b UT322 and UT325 Keypad
Press to start or exit Setup.
SETUP
Press to scroll the Setup option you want to change
(See "Changing Setup Options")
After entering the Setup mode, press to increase the displayed setting.
(See "Changing Setup Options")
After entering the Setup mode, press to decrease the displayed setting.
(See "Changing Setup Options")
Confirm button.
ENTER
(See "Changing Setup Options")
Using the Thermometer
1) Plug the thermocouple(s) into the input connector(s).
2) Press the power button to turn on the thermometer.
3) Set the type of thermocouple(s) to be the same with the one plugged into the
input connector(s).
If no thermocouple is plugged into the selected input terminal or the thermocouple
is "open" or the over-range positive deviation is too big, the display
shows "_ _ _ _ ".
Changing Setup Options
Use Setup to change the following settings:
Logging interval
Thermocouple type
Offset (T1)
UT325 only)
Sleep Mode
Line Frequency
Time (S-T)
(Lo) (UT323 and UT325 only)
High Limit Alarm (Hi) (UT323 and UT325 only)
Over Limit Signal Output (SI) ON/OFF
(UT323 and UT325 only) Normal
Temperature Compensation (NTC) ON/OFF
DEBUG ON/OFF (UT323 and
UT325 only)
Save setting and return to normal measurement mode.
When the thermometer is in Setup mode, the display shows and blinks SETUP.
Press SETUP to start or exit Setup.
Press SETUP to scroll through the Setup options you want to change.
The logging interval determines how often the thermocouple stores logged
readings in memory. You choose the length of the logging interval. See "Using
Memory".
Press SETUP until the display shows INTERVAL.
The thermometer stores logged readings at the end of each logging interval.
You can select a logging interval by pressing
or . Holding down or
the number to change more quickly. The maximum interval is 59:59 and the
minimum interval is 00:00. When the logging interval is 00.00, you need to store
the readings manually as the auto store feature will be disabled.
1. Press SETUP until the display shows TYPE.
2. Press
or
to select the thermocouple type you want including K-, J-, E- type
(UT323 and UT325 have extra R-, S- and N- ttype, UT321/UT325 can use
TYPE button to change directly)).
1. Press SETUP until the display shows OFFSET and T1
2. Press
or
to change the offset value. The offset range is -6~6.
1. Press SETUP until the display shows OFFSET and T2
2. Press
or
to change the offset value. The offset range is -6~6.
to change until the desired sleep time is obtained.
toggle between time format "h:m" and "m:s".
5. If you don't change this option, it means the time that the meter has run
through currently.
6. The system counts the time when the user turns on the thermometer. It will
be automatically cleared once the Thermometer is dis-energized.
6. The Min.value for low alarm is the lower limit value of measuring range of the
applied thermocouple.
7. The Max. value for low alarm is the setting high alarm limit value if high alarm
is enabled (if still higher than high alarm limit value, Max. value=High Alarm
Limit Value-1), otherwise this Max. value is the upper limit value of measuring
range of the applied thermocouple.
Min. value
Max. value
Offset (T2) (UT322 and
Low Limit Alarm
causes
High Alarm Limit Value-1

Advertisement

loading

Summary of Contents for UNI-T UT323

  • Page 1 UT322: dual input(T1,T2), suitable for K-, J-, T- and E- type thermocouples When the thermometer is in Setup mode, the display shows and blinks SETUP. UT323: single input(T1), suitable for K-, J-, T-, E-, R-, S- and N- type thermocouples. Press SETUP to start or exit Setup.
  • Page 2 Setup display 1 shows the logged reading. Each presses of ENTER will automatically defaults to OFF status. UT325 only), Over Limit Signal Output (UT323 and store the logged readings in the next memory location. Press could UT325 only), Normal Temperature Compensation, change the memory location, the empty location shows “- - - - - ”.