Summary of Contents for HT optec HT210

  • Page 1 ENGLISH User manual  Copyright HT ITALIA 2017 Release EN 2.00 - 01/09/2017...
  • Page 2: Table Of Contents

    HT210 Table of content: SAFETY PRECAUTIONS AND PROCEDURES ............2     1.1.  Preliminary instructions ..................... 2  1.2.  During use ......................... 3  1.3.  After use ..........................3  1.4.  Measuring (overvoltage) categories definitions ..............3    GENERAL DESCRIPTION ................... 4    ...
  • Page 3: Safety Precautions And Procedures

    HT210 1. SAFETY PRECAUTIONS AND PROCEDURES This meter is compliance with safety Standard IEC/EN61010-1 related to electronic measuring instruments. For your own safety and to avoid damaging the instrument follow the procedures described in this instruction manual and read carefully all notes preceded by this symbol .
  • Page 4: During Use

    HT210 1.2. DURING USE Read the recommendations which follow and the instructions in this manual: CAUTION An improper use may damage the instrument and/or its components or injure the operator.  When changing the range, first disconnect the test leads from the circuit under test in order to avoid any accident ...
  • Page 5: General Description

    HT210 2. GENERAL DESCRIPTION The instrument performs, in completely Autorange, the below measurements:  DC Voltage  AC Voltage  Resistance and Continuity test  Capacitance  Frequency  Duty cycle  Diode test  Temperature with type K probe All selectable by means of a 5 position rotary selector (including OFF position).
  • Page 6: Nomenclature

    HT210 4. NOMENCLATURE 4.1. INSTRUMENT DESCRIPTION CAPTION: 1. LCD display 2. MODE key 3. REL key 4. Hz% key 5. HOLD 6. Function selector 7. COM input jack VHz% input jack Fig. 1: Instrument description EN - 5...
  • Page 7: Description Of Function Keys

    HT210 4.2. DESCRIPTION OF FUNCTION KEYS 4.2.1. HOLD By pressing HOLD key the measured value is frozen on the display where the symbol "HOLD" appears. Pressing again HOLD to disable this function and resume normal operation. By pressing the key at least 1 second the enable/disable of backlight feature can be performed.
  • Page 8: Operating Instructions

    HT210 5. OPERATING INSTRUCTIONS 5.1. DC VOLTAGE MEASUREMENT CAUTION The maximum input for DC voltage is 600V. Do not attempt to measure higher voltages to avoid electrical shocks or damages to the instrument. Fig. 2: Using the meter for DC Voltage measurement 1.
  • Page 9: Ac Voltage Measurement

    HT210 5.2. AC VOLTAGE MEASUREMENT CAUTION The maximum input for AC voltage is 600V. Do not attempt to measure higher voltages to avoid electrical shocks or damages to the instrument. Fig. 3: Using the instrument for AC Voltage measurement 1. Switch on .
  • Page 10: Resistance And Continuity Test Measurements

    HT210 5.3. RESISTANCE AND CONTINUITY TEST MEASUREMENTS CAUTION Before taking resistance measurements in circuit remove power from the circuit being tested and discharge all capacitors. Fig. 4: Using the instrument for Resistance and Continuity test measurements 1. Switch on  .
  • Page 11: Diode Test

    HT210 5.4. DIODE TEST CAUTION Before taking diode test measurements in circuit remove power from the circuit being tested and discharge all capacitors. Fig. 5: Using the instrument for Diode test 1. Switch on  2. Pressing the MODE key until the symbol is shown at display to meaning the diode test 3.
  • Page 12: Frequency And Duty Cycle Measurements

    HT210 5.5. FREQUENCY AND DUTY CYCLE MEASUREMENTS CAUTION The maximum input for AC voltage is 250V. Do not attempt to measure higher voltages to avoid electrical shocks or damages to the instrument. Fig. 6: Using the instrument for Frequency and Duty cycle measurements 1.
  • Page 13: Capacitance Measurement

    HT210 5.6. CAPACITANCE MEASUREMENT CAUTION Before taking capacitance measurement in circuit remove power from the circuit being tested and discharge all capacitors. Connect the test capacitor to the inputs respecting the polarity connections when required. Fig. 7: Using the instrument for Capacitance measurement 1.
  • Page 14: Temperature With Type K Probe Measurement

    HT210 5.7. TEMPERATURE WITH TYPE K PROBE MEASUREMENT CAUTION Before taking temperature measurements in circuit remove power from the circuit being tested and discharge all capacitors. Fig. 8: Using the instrument for Temperature with type K probe measurement 1. Switch on °C or °F positions 2.
  • Page 15: Maintenance

    HT210 6. MAINTENANCE 6.1. GENERAL INFORMATION  In order to go guarantee the performances of the instrument be sure to use it according to these instructions and keep it stored on suitable environmental conditions  Do not expose it to high temperatures or humidity or direct sunlight ...
  • Page 16: Technical Specifications

    HT210 7. TECHNICAL SPECIFICATIONS 7.1. TECHNICAL FEATURES Accuracy calculated as [%reading + (num. dgts*resolution)] at 18°C28°C, <70%RH DC Voltage (Autorange) Range Resolution Accuracy Input impedance Overload protection 4.000V 0.001V (1.2%rdg+2dgt) 40.00V 0.01V 600VDC/ACrms 7.8M 400.0V 0.1V (1.5%rdg+2dgt) 600V AC Voltage (Autorange) Accuracy Range Resolution...
  • Page 17: Reference Guidelines

    HT210 Capacitance (Autorange) Range Resolution Accuracy Overload protection (5.0%rdg + 7dgt) 40.00nF 0.01nF 400.0nF 0.1nF (3.0%rdg + 5dgt) 4.000F 0.001F 250VDC/ACrms 40.00F 0.01F (5.0%rdg +5dgt) 100.0F 0.1F Temperature with type K probe (Autorange) Range Resolution Accuracy Overload protection -20°C ÷ 400°C 0.1°C (3.0%rdg + 5°C) 400°C ÷...
  • Page 18: Accessories

    HT210 7.3. ACCESSORIES 7.3.1. Accessories provided  Pair of test leads  Adapter + K-type wire probe  Battery  Carrying bag  User manual 7.3.2. Optional accessories  Type K probe for air and gas Cod. TK107  Type K probe for semi solid substances Cod.
  • Page 19: Service

    HT210 8. SERVICE 8.1. WARRANTY CONDITIONS This instrument is guaranteed against material or production defects, in accordance with our general sales conditions. During the warranty period the manufacturer reserves the right to decide either to repair or replace the product. Should you need for any reason to return back the instrument for repair or replacement take prior agreements with the local distributor from whom you bought it.