Fluke 52120A-C User Manual page 89

Transconductance amplifier
Table of Contents

Advertisement

and negative values as given above.
52. Send the command: :CAL:ADJ? GAIN,20.0000,<calculated value>
53. Push S on the 5720A.
54. Push A on the Product.
55. Set the 5720A output to the values in Table 5-7 for the 20 A shunt and 200 mA dc
and -200 mA dc adjustment points.
56. Push M on the 5720A.
57. Push M on the Product.
58. Write the values shown on the DMM in the Measured Current column of Table 5-7.
Calculate the average amplitude of the absolute positive and negative values as
given above.
59. Send the command: :CAL:ADJ? GAIN,20.0000,<calculated value>
60. Push S on the 5720A
61. Push V on the Product.
62. Set the DMM to the ACV function and the 2V range.
63. Set the 5720A voltage output and frequency to the values in Table 5-7 for the 20 A
shunt and 2 V at 6 kHz adjustment point.
64. Push M on the 5720A.
65. Push M on the Product.
66. Write the value shown in the DMM in the Measured Current column of Table 5-7.
67. Send the command: :CAL:ADJ? FLAT,20.0000,<measured value>
68. Push S on the Product and the 5720A.
69. Set the HIGH terminals on the Product.
70. Connect the 100 A shunt between the Product output and the input of the DMM. See
Figure 5-2.
71. Set the DMM to the DCV function on the 1V range.
72. Set the DMM to Resolution 6.
73. Turn off all DMM Math functions.
74. After the DMM measurement settles, push the Offset key.
The DMM must show a measurement between 0.000008 V and -0.000008 V.
75. Write in the DMM measurement in the measured error column in the Output Off
and 100 A shunt row of Table 5-7.
76. Set the 5720A output to the values in Table 5-7 for the 100 A shunt and 0 V dc
adjustment points.
77. Push M on the 5720A.
78. Push D on the Product.
79. Push V on the Product.
80. Push M on the Product.
Note
5
Calibration
Introduction
5-19