Introduction; Notes On Product Use; Accuracy - Fluke 52120A-C User Manual

Transconductance amplifier
Table of Contents

Advertisement

Introduction

The Product can be configured to operate in a stand-alone configuration or controlled by
an Electrical Power Standard like the Fluke 6100B or 6105A. The inputs of multiple
Products can be connected to one power standard. Then the current output from each
Product can be connected in parallel to a load. Each Product can be set so that each
output has a different ac phase angle.

Notes on Product Use

Accuracy

The specifications in Chapter 1 for stand-alone operation do not include the errors of the
instrument that supplies the voltage or current signal to the Product input. To get the
absolute accuracy of the current output, use the "root sum of squares" (RSS) procedure.
To calculate total accuracy:
First calculate:
√ specA
Multiply the total specification by the coverage factor to get the expanded accuracy at the
necessary confidence probability. The most commonly used confidence probabilities for
this type of product are shown in Table 2-1.
Confidence Probability
In this example, the 5720A has an output of 1.2 volts at 20 Hz applied to the Product. The
Product is set to the 120 A range with LCOMP off. The Product and 5720A have
confidence probability specifications of 99 %. It is not necessary to change them to
standard confidence before you calculate the specification with the RSS procedure. For
this example, the 5720A has a confidence probability of 95 % and the Product 99 %.
Table 2-2 shows how the total accuracy is calculated.
5720A output of 1.2 V at 20 Hz (240 ppm + 40 μV)
Divisor to convert to standard confidence
5720a standard confidence
Product output 120 Amp at 20 Hz, LCOMP off
Divisor to convert to standard confidence
Product standard confidence (0.015 % + 0.020 %)
Root sum of squares (RSS)
Multiplier to convert to 99 % confidence
Combined accuracy at 99 % confidence
2
2
+ specB
= total specification
Table 2-1. Confidence Probabilities
95 %
99 %
Not Stated
Table 2-2. Total Accuracy Calculation
Calculation Steps
Front-Panel Operation
Coverage Factor Divisors/Multipliers
2.00
2.58
1.73
Expanded
0.0273 %
2.00
0.0350 %
2.58
2.58
0.0497 %
2

Introduction

Standard
0.0137 %
0.0136 %
0.0192 %
2-3