® PATIENT CHAIR CARRIAGE ADJUSTMENT YB-782 Version 1.1...
Page 2
If the tolerances are too small the chair carriage can bind on the chair guide causing noises and the lifting motor to be over-worked with risk for the motor to overheat and cause a decrease in lifespan. Adjustment Sequence Version 1.10 YB-782...
Page 3
Tighten the nut while keeping the bolt in position. The bottom half of the chair carriage can move a lot during this adjustment. Both sides of the lower part of the carriage should have slight contact with the chair guide now. Version 1.10 YB-782...
Page 4
2 tracks in the grease on each side of the chair guide, see picture above. If there is only 3 or less tracks in the grease, a re-adjustment is needed. Check that both rollers still have contact with the chair guide. Version 1.10 YB-782...
Page 5