General Operations; Connecting To The Readout Device; Drive Current; Stability Of Readings - AccuMac AM1751 User Manual

Platinum resistance thermometer
Table of Contents

Advertisement

AM1751 Secondary Reference PRT User's Guide

3 General Operations

3.1 Connecting to the readout device

The AM1751 is equipped with a four-wire cable (see Figure 1). Four
lead wires are used to cancel lead wire resistance. For best results, the
readout device should be equipped to handle four-terminal resistors.
The lead wires can be distinguished by insulation colors. Lead wire
pairs attached to each end of the sensor are identified by red/black and
white/blue insulation.
White
Blue

3.2 Drive Current

AccuMac recommends 1mA as drive current to ensure the best
measurement.

3.3 Stability of Readings

To achieve the best accuracy, allow sufficient time for PRT to
stabilize before taking the readings.

3.4 Immersion Requirements

Stem effect can cause measurement errors due to heat lost or gained
by the sensing element through the thermometer stem. To minimize
the error, appropriate immersion depths are required. A practical way
to determine the minimum immersion depths is to change the depth
Rev. 201609V3
Figure1
Red
Black
5/10

Advertisement

Table of Contents
loading

Table of Contents