Sensorex SD7500CD Product Instruction Sheet

Differential ph/orp electrode

Advertisement

Quick Links

SD7000 Series Differential
pH/ORP Electrode Instructions
Introduction
Congratulations on your purchase of your Sensorex SD7000
Series pH or ORP true differential electrode. Please follow
instructions and tips to get the maximum life out of your
electrode.
Electrode Care and Use Tips
1. The SD7000 Series pH and ORP electrodes are shipped with a
cap containing a solution of pH 4 buffer and potassium
chloride.
The cap should remain on the electrode until it is used. If the
electrode is used infrequently the cap and its solution should
be saved and the electrode stored in it. Store pH and ORP
electrodes in pH 4 buffer if soaking solution is discarded.
Before using the electrode remove the soaker cap as shown
in FIG 1.
2. Sensorex true differential electrodes have a 2 year prorated
warranty from the date etched in the electrode body as shown
in FIG 2. Electrodes in inventory should be rotated so that the
older electrodes are used first. The date code is stamped on
each electrode by fiscal week and year (Example: 4613 = week 2
of November 2013).
3. After exposure to a sample, buffer or rinse solution, carryover
can be minimized by blotting—never by wiping—the elec-
trode with a clean non-abrasive paper or a clean cloth towel.
Do not use a brush on pH glass.
4. As a rinse solution, measure and use a portion of the next
sample or buffer. This action will minimize carryover
contamination.
5. When calibrating pH electrodes, use a buffer close in value to
that expected from the sample for one-point calibrations or as
the second buffer for two-point calibrations (See below). This
action will minimize span errors. ORP electrodes are typically
not calibrated. However, a solution to verify electrode perfor-
mance should be used periodically. 225mV solution, Zobell's
solution (Part no. B225), is a good choice for single point
verification.
Parts covered by this product data sheet include:
SD7000CD, SD7000CD-ORP, SD7500CD, SD7500CD-ORP, SD7420CD, SD7420CD-ORP
Page 1 of 8
PRODUCT INSTRUCTION SHEET
FIG. 1
FIG. 2
Form: InstructSD7000CD-06 [Rev: 2016-10-10]

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SD7500CD and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Sensorex SD7500CD

  • Page 1 Before using the electrode remove the soaker cap as shown in FIG 1. 2. Sensorex true differential electrodes have a 2 year prorated warranty from the date etched in the electrode body as shown in FIG 2. Electrodes in inventory should be rotated so that the older electrodes are used first.
  • Page 2: Mechanical Installation

    PRODUCT INSTRUCTION SHEET 6. Readings stabilize faster in some solutions than in others; allow FIG. 3 time for the readings to stabilize. In general, new pH electrodes stable readings display in 10-15 seconds. ORP stability in Conduit box samples can take much longer. 7.
  • Page 3 PRODUCT INSTRUCTION SHEET Salt Bridge Removal and Replacement FIG. 6 FIG. 5 1. Remove the salt bridge using an adjustable wrench or pliers O-ring by turning counterclockwise. SEE FIG 5. must be O-ring 2. Pull up on salt bridge to remove. SEE FIG 6. seated in must be groove...
  • Page 4 PRODUCT INSTRUCTION SHEET Electrode Calibration Guidlines: FIG. 8A FIG. 8B As a rule, follow the procedures shown in the pH Meter's Instruction Manual. These procedures will vary depending on whether the meter is a simple type with manual adjustments, a micro-processor type or a pH transmitter.
  • Page 5 These sensors are loop powered 4-20mA. 12-30V DC Shark power is required (>24V DC is recommended). TX/P Transmitter Figure 19 is wiring for SD7500CD & SD7500CD-ORP sensors. These sensors DO NOT REQUIRE POWER. Do not connect to power. These are self-powered. FIG. 12 GLI PRO P3...
  • Page 6 The resistance should be between 250 and 350 ohms at approximately 25 ºC (SD7000CD only). For SD7500CD the resistance should measure about 1100 ohms at 25 ºC. If the temperature element is good, reconnect the yellow and black wires to the module.
  • Page 7: Product Specifications

    Titanium (ground pin), Viton (o-rings) Cable Length: 20ft Max flow Rate: 10ft./sec Max transmission distance: 3000ft (914m) SD7500CD , SD7500CD-ORP pH Range: 0-14 ORP Range: -2000 to + 2000mV Temp/Pressure: Min 32° F(0°C)/ Max 185° F/(85° C) Speed of Response: 95% in 5 seconds...
  • Page 8: Limited Warranty

    20ft cable, tinned leads, 1.5” NPT mounting, self- powered. Replacement salt bridges sold separately. FIG. 19 SD7500CD -ORP Differential ORP sensor with replaceable salt bridge, 20ft cable, tinned leads, 1.5” NPT mounting, self- powered. Replacement salt bridges sold separately.

This manual is also suitable for:

Sd7000cdSd7000cd-orpSd7000 seriesSd7500cd-orpSd7420cdSd742cd-orp

Table of Contents