SD7000 Series Differential
pH/ORP Electrode Instructions

Introduction

Congratulations on your purchase of your Sensorex SD7000
Series pH or ORP true differential electrode. Please follow
instructions and tips to get the maximum life out of your
electrode.

Electrode Care and Use Tips

1. The SD7000 Series pH and ORP electrodes are shipped with a
cap containing a solution of pH 4 buffer and potassium
chloride.
The cap should remain on the electrode until it is used. If the
electrode is used infrequently the cap and its solution should
be saved and the electrode stored in it. Store pH and ORP
electrodes in pH 4 buffer if soaking solution is discarded.
Before using the electrode remove the soaker cap as shown
in FIG 1.
2. Sensorex true differential electrodes have a 2 year prorated
warranty from the date etched in the electrode body as shown
in FIG 2. Electrodes in inventory should be rotated so that the
older electrodes are used first. The date code is stamped on
each electrode by fiscal week and year (Example: 4613 = week 2
of November 2013).
3. After exposure to a sample, buffer or rinse solution, carryover
can be minimized by blotting—never by wiping—the elec-
trode with a clean non-abrasive paper or a clean cloth towel.
Do not use a brush on pH glass.
4. As a rinse solution, measure and use a portion of the next
sample or buffer. This action will minimize carryover
contamination.
5. When calibrating pH electrodes, use a buffer close in value to
that expected from the sample for one-point calibrations or as
the second buffer for two-point calibrations (See below). This
action will minimize span errors. ORP electrodes are typically
not calibrated. However, a solution to verify electrode perfor-
mance should be used periodically. 225mV solution, Zobell's
solution (Part no. B225), is a good choice for single point
verification.
Parts covered by this product data sheet include:
SD7000CD, SD7000CD-ORP, SD7500CD, SD7500CD-ORP, SD7420CD, SD7420CD-ORP
Page 1 of 8
PRODUCT INSTRUCTION SHEET
FIG. 1
FIG. 2
Form: InstrSD7000-05 [Rev: 2016-08-15]
©2016 Sensorex Corporation

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SD7000 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Sensorex SD7000 Series

  • Page 1 Electrode Care and Use Tips 1. The SD7000 Series pH and ORP electrodes are shipped with a cap containing a solution of pH 4 buffer and potassium chloride.
  • Page 2 In-Line Installation Use pipe sealing tape or paste on the 1.5" threads near the pH glass and salt bridge. Thread into user-supplied SCH80 1.5" threaded tee. See FIG 4. Page 2 of 8 Form: InstrSD7000-05 [Rev: 2016-08-15] ©2016 Sensorex Corporation...
  • Page 3 Protein-based coatings are best removed with an enzyme- based cleaner such as TERG-A-ZYME (www.alconox.com). Abrading or sanding a pH electrode’s surface should never be done. Page 3 of 8 Form: InstrSD7000-05 [Rev: 2016-08-15] ©2016 Sensorex Corporation...
  • Page 4 It is important that the sample being measured by the lab meter be at the process temperature or erroneous results may occur. Page 4 of 8 Form: InstrSD7000-05 [Rev: 2016-08-15] ©2016 Sensorex Corporation...
  • Page 5 SD7000CD and SD7000CD-ORP sensors only. If you don't see a wiring diagram for your make and model number of controller, please contact Sensorex Technical support at tech- support@sensorex.com or call 714-895-4344. Figure 18 is wiring for the SD7420CD & SD7420CD-ORP sen- Aquametrix FIG.
  • Page 6 PRODUCT INSTRUCTION SHEET Troubleshooting Your SD7000 Series Electrode FIG. 13 GLI P63 General Troubleshooting Always check all electrical connections. Make sure all parts are assembled correctly and o-rings are well greased. Electrode Troubleshooting 1. Put the sensor in a pH 7 buffer solution and wait for the tempera- ture of the sensor and buffer to reach room temperature.
  • Page 7 Standard Refill solution for PH and ORP differential sensors, 150mL SDS-7025 Standard Refill solution for PH and ORP differential sensors, 250mL SDS-7050 Standard Refill solution for SD7000 Series sensors, 500mL Page 7 of 8 Form: InstrSD7000-05 [Rev: 2016-08-15] ©2016 Sensorex Corporation...
  • Page 8 There are no implied warranties of merchantability or fitness for a particular purpose given in connection with the sale of any goods. In no event shall SENSOREX be liable for conse- GRN CLR WHT YEL BLK quential, incidental or special damages.

Table of Contents