Sensorex SD7000CD Series Product Instruction Sheet

Differential ph/orp electrode

Advertisement

SD7000CD Series Differential
pH/ORP Electrode Instructions
Introduction
Congratulations on your purchase of your Sensorex SD7000CD
pH or ORP true differential electrode. Please follow instruc-
tions and tips to get the maximum life out of your electrode.
Electrode Care and Use Tips
1. The SD7000CD pH and ORP electrodes are shipped with a cap
containing a solution of pH 4 buffer and potassium chloride.
The cap should remain on the electrode until it is used. If the
electrode is used infrequently the cap and its solution should
be saved and the electrode stored in it. Store pH and ORP
electrodes in pH 4 buffer if soaking solution is discarded.
Before using the electrode remove the soaker cap as shown
in FIG 1.
2. Sensorex true differential electrodes have a 1 year warranty
from the date etched in the electrode body as shown in FIG 2.
Electrodes in inventory should be rotated so that the older
electrodes are used first. The date code is stamped on each
electrode by fiscal week and year (Example: 4613 = week 2 of
November 2013).
3. After exposure to a sample, buffer or rinse solution, carryover
can be minimized by blotting—never by wiping—the elec-
trode with a clean non-abrasive paper or a clean cloth towel.
Do not use a brush on pH glass.
4. As a rinse solution, measure and use a portion of the next
sample or buffer. This action will minimize carryover
contamination.
5. When calibrating pH electrodes, use a buffer close in value to
that expected from the sample for one-point calibrations or as
the second buffer for two-point calibrations (See below). This
action will minimize span errors. ORP electrodes are typically
not calibrated. However, a solution to verify electrode perfor-
mance should be used periodically. 225mV solution, Zobell's
solution (Part no. B225), is a good choice for single point
verification.
Parts covered by this product data sheet include:
SD7000CD, SD7000CD-ORP
Page 1 of 8
PRODUCT INSTRUCTION SHEET
FIG. 1
FIG. 2
Form: InstrSD7000-04B [Rev: 2015-12-14]
©2014 Sensorex Corporation

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SD7000CD Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Sensorex SD7000CD Series

  • Page 1 Before using the electrode remove the soaker cap as shown in FIG 1. 2. Sensorex true differential electrodes have a 1 year warranty from the date etched in the electrode body as shown in FIG 2. Electrodes in inventory should be rotated so that the older electrodes are used first.
  • Page 2: Mechanical Installation

    In-Line Installation Use pipe sealing tape or paste on the 1.5" threads near the pH glass and salt bridge. Thread into user-supplied SCH80 1.5" threaded tee. See FIG 4. Page 2 of 8 Form: InstrSD7000-04B [Rev: 2015-12-14] ©2014 Sensorex Corporation...
  • Page 3 Protein-based coatings are best removed with an enzyme- based cleaner such as TERG-A-ZYME (www.alconox.com). Abrading or sanding a pH electrode’s surface should never be done. Page 3 of 8 Form: InstrSD7000-04B [Rev: 2015-12-14] ©2014 Sensorex Corporation...
  • Page 4 It is important that the sample being measured by the lab meter be at the process temperature or erroneous results may occur. Page 4 of 8 Form: InstrSD7000-04B [Rev: 2015-12-14] ©2014 Sensorex Corporation...
  • Page 5: Product Specifications

    ORP controller or transmitter. If you dont't see a wir- distance: 3000ft (914m) ing diagram for your make and model number of controller, please contact Sensorex Technical support at techsupport@ sensorex.com or call 714-895-4344. Aquametrix FIG. 10...
  • Page 6 Possible Cause Corrective Action a) Cracked pH glass a) Replace electrode 6.2-6.8 in all buffers b) Stress crack b) Contact Sensorex for Return Authorization a) Bad connection a) Check/fix connection 7.00 in all buffers b) Internal short circuit b) Contact Sensorex for...
  • Page 7: Ordering Information

    In no event shall SENSOREX be liable for consequential, incidental or special damages. The buyer’s sole and exclusive remedy and the limit of SENSOREX’s liability for any loss whatsoever shall...
  • Page 8 PRODUCT INSTRUCTION SHEET NOTES: FIG. 17 HACH SC200 pH/ORP/DO Module GRN CLR WHT YEL BLK Page 8 of 8 Form: InstrSD7000-04B [Rev: 2015-12-14] ©2014 Sensorex Corporation...