Wavecrest DTS-2070 Calibration Procedure page 9

Advertisement

WAVECREST Corporation
8.0
Arming and Gating Checks
(8-1) Connect CAL2 to the ARM1 input channel. Connect CAL1 to the delay line and the
delay line to the CH1 input channel.
(8-2) At Menu #7, select PERIOD, CH1, TRIG = 50%-50%. Select AUTO ARM and
CONTINUOUS mode on the front panel, and press FUNC. Verify the displayed
measurement is 5000 psec ±10 psec with JITTER <6 psec.
(8-3) Set the function generator to output a square wave with a 5 nsec wide duty cycle, and a
1V
pk-pk
(8-4) Select EXTERNAL ARMING on ARM2 mode. Verify sample size = 10,000.
(8-5) Connect the CAL1 to the CH1 input channel and CAL2 to the CH2 input channel.
Connect the output of the function generator to the ARM2 input. Go to Menu #9 and set
GATE = "ON H". Then press CONTINUOUS.
(8-6) Determine the approximate amount of time it takes the DTS to obtain l0,000 samples.
This can be done by watching Menu #6. Then change the gating duty cycle on the
function generator from 5 nsec to 2 nsec. The time to obtain the 10,000 samples should
now be more than twice as long.
(8-7) Remove all test equipment and cabling from the DTS.
9.0
Shut Down the DTS System
(9-1) Go to Menu #1 and select CLEAR. Then press FUNC to return the system to the default
state.
(9-2) Power off the standby switch on the front panel and the main power switch on the rear of
the instrument. Remove all cables and unplug the AC power cord from the power source
and package the accessories that accompanied the system for return to the customer.
9.1
End of Calibration Procedure
If any display function does not appear to provide reasonable results, notify the
WAVECREST Corporation factory for support at 1-800-733-7128.
DTS-2070C Calibration Procedure
signal at 1 KHz.
7
(17 April 1998)