5.2

SYSTEM ON

Activate the power to the system by rotating the "Main Breaker" handle to
ON. If the display on the control panel does not illuminate, ensure that
the EMERGENCY OFF button is in the operate position. Press the reset
button to engage control power.
5.3

HEATER MODULE POWER

When the RESET button is pressed, the power contactors will engage.
5.4

PROCESS ALARMS

If a fault occurs, the main contactor will disengage, an indicator LED will
turn red to show the type of fault, and an audible alarm will sound.
Pressing the RESET button will silence the audible alarm. If the RESET
button is pushed and there are no active faults, the heater will restart.
PAGE 32
QUANTUM DI HEATER OPERATION / MAINTENANCE MANUAL