SAALMANN Saalux User Manual page 25

Uv comb
Table of Contents

Advertisement

IEC 61000-4-8
NOTE:
U
is the a. c. mains voltage prior to application of the test level.
T
®
The Saalux
device is designed for operation in the electromagnetic environment specified below. The
user should ensure that the device is used in such an environment.
Immunity test
IEC 60601-
test level
Conducted RF
3 V rms
IEC 61000-4-6
150 kHz to 80 MHz
6 V rms
150 kHz to 80 MHz
within ISM bands
and amateur radio
bands
Radiated
10 V/m
RF
80 MHz to 2700
IEC 61000-4-3
MHz
according to IEC
Radiated
60601-1-2:2014
RF according
Table 9
IEC 61000-4-3
in close
proximity to
wireless
communication
devices
NOTE: These guidelines may not apply in all situations. Electromagnetic is affected by absorption and
reflection from structures, objects and people.
®
User manual Saalux
UV comb
Electromagnetic Immunity
Electromagnetic environment – guidance
Compliance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the Saalux
device, including cables, than the recommended
separation distance (d in meter) calculated from the
equation applicable to the frequency of the
transmitter.
3 V rms
6 V rms
10 V/m
passed
(corresponds to a recommended safety distance of
0.3 m to the devices of the corresponding radio
services)
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range. Interference may occur in the
vicinity of equipment marked with the following
symbol
®
page 23