Appendix 1 – Electromagnetic compatibility
®
The Saalux
device is designed for operation in the electromagnetic environment specified below. The
user should ensure that the device is used in such an environment.
interference emission
RF emission
according CISPR 11
RF emission
according CISPR 11
Harmonic emission according
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
®
The Saalux
device is designed for operation in the electromagnetic environment specified below. The
user should ensure that the device is used in such an environment.
Immunity test
IEC 60601-
Electrostatic
± 8 kV
discharge
contact discharge
(ESD)
max. ± 15 kV
IEC 61000-4-2
air discharge
Electrical
± 2 kV
transients /
for power supplies
bursts
IEC 61000-4-4
Surge
± 1 kV for line(s)
to line(s)
IEC 61000-4-5
Voltage dips,
0 % U
short
for 0.5 cycle
interruptions
0 % U
and voltage
for 1 cycle
variations on
70 %
power supply
for 25 cycles
input lines
0 % U
IEC 61000-4-11
for 250/300 cycles
Power
30 A/m
frequency
(50/60 Hz)
magnetic field
®
User manual Saalux
UV comb
Electromagnetic emissions
Compliance
Group 1
Class B
Not applicable
Complies
Electromagnetic immunity
Compliance
test level
level
± 8 kV
max. ± 15 kV
± 2 kV
± 1 kV
0 % U
T
T
for 0.5 cycle
0 % U
T
T
for 1 cycle
70 %
for 25 cycles
0 % U
T
T
for 250/300 cycles
30 A/m
Electromagnetic surrounding - Guidance
®
The Saalux
device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
®
The Saalux
device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Electromagnetic environment –
guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Saalux
continued operation during power mains
interruptions, it is recommended that the
®
Saalux
device is powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
®