How to Apply Electrical Stimulation (Figure 3)
1. Turn on the power.
- Rotate the Intensity Dial/Power Button downwards. When you hear a click, the power is turned
on and the stimulation starts.
- The special mode of KNEAD, REPEAT and WIDE has been pre-set by the research team.
These parameters provide 25 Hz pulses with 4-second stimulation and 1-second pause cycle.
- Then you will see the Massage Button at KNEAD and REPEAT lights are flashing. At the unit
screen, counting down 15 minutes, frequency bars and WIDE will be displayed.
Figure 3
2. Select one stimulation method.
- If you could not see the setting as mentioned above or if you noticed a different flashing
location, you can reset it by pressing the button of MASSAGE until the green light stops at
KNEAD, press REPEAT button to see its green light turned on and press WIDE button to see
WIDE on the screen.
- When you use the unit for the second time and after, the Buttons and settings you used in the
previous setup will flash and automatically start.
- If the Pads come off from the skin, the power will be turned off automatically in approximately 7
seconds. If it happens within the first half of stimulation, you can re-attach the Pad and restart
3