A&D UA-651BLE Instruction Manual page 29

Upper arm
Hide thumbs Also See for UA-651BLE:
Table of Contents

Advertisement

Available languages

Available languages

Technical Data
Bluetooth
devices carry the Bluetooth
®
iPhone, iPad and iPod are trademarks of Apple Inc., registered in the
U.S. and other countries.
Google Play and Google Play logo are registered trademarks of Google,Inc.
EMD Technical Data
Battery-operated or AC Adapter-operated Blood Pressure Monitor
Medical Electrical Equipment needs special precautions regarding
EMD and needs to be installed and put into service according to the
EMD information provided in the following.
Portable and mobile RF communication equipment (e.g. cell phones)
can affect Medical Electrical Equipment.
The use of accessories and cables other than those specified may
result in increased emissions or decreased immunity of the unit.
Table 1 - EMISSION Limits -
Phenomenon
Conducted and radiated RF EMISSION
CISPR 11
Harmonic distortion IEC 61000-3-2
Voltage fluctuations and flicker
IEC 61000-3-3
Table 2 - IMMUNITY TEST LEVELS : Enclosure Port -
Phenomenon
Electrostatic discharge IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
Rated power frequency magnetic fields
IEC 61000-4-8
logo mark.
®
Compliance
Group 1, Class B
Class A
Compliance
IMMUNITY TEST LEVELS
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
See table 4
30 A/m
50 Hz or 60 Hz
E-28