Download Print this page

Advertisement

Quick Links

Operations Manual
Item: R4040 and R4045 (-E)
I. Specifications:
Speed:
Adjustable from 1 to 99, (1 to 65rpm)*
Capacity: R4040: 6 x 50ml R4045: 12 x 50ml
Timer: 1min to 99hr 59 min / Cont.
Dimensions: (wxdxh)
R4040: 29 x 13.7 x 19.9 cm (11.4 x 5.4 x 7.8 in.)
R4045: 41 x 13.7 x 19.9 cm (16.1 x 5.4 x 7.8 in.)
Weight: R4040: 4lbs / 1.8kg
R4045: 5lbs / 2.27kg
Electrical: 100 to 240VAC, 50-60Hz, 3A
Operating environment: 4°C (cold room) to 70°C
Warranty: 2 years
*Maximum speed may be limited in modes with a mixing function.
II. Product Set-Up:
Place the rotator on a clean and stable surface.
Ensure that there is at least 3 inches of open space
both in front of and behind the instrument.
V1.0
Plug the instrument into an outlet of the appropriate
voltage (100 to 240V). Press and hold the power
button for 3 seconds, the instrument powers on.
III. Tube Holder Installation / Exchange:
The instrument includes an assortment of tube
holders. To exchange the tube holders, remove the
rotational bar from the instrument by unscrewing
the thumb screw on the left side.
loosened the bar can now be removed by pressing it
fully to the right and then lifting the left side
upward. Tube holders can be installed or removed
from the rotational bar by sliding them in either the
left or right direction.
To load test-tubes, press them firmly into the
appropriate holder. ALWAYS ensure a balanced load
across both sides of the rotational axis.
mixing 15 and 50ml tubes, always load a similar
quantity of tubes on each side of the rotational axis.)
IV. Product Operation:
Choose the Mixing Mode: Press the "Mode" to
toggle through the available modes:
F1 – F9
Rotational Modes
C1-C2
Custom Modes
-- --
Short Rotational Mode (aggressive mixing)
Set the Rotational Speed:
After choosing the mode, the rotational speed can
be selected by pressing the up and down keys to
adjust the display to the desired speed setting.
(Level: 1 to 99)
Set the Run Time and Frequency:
Press the "SET" key so that the run time is flashing.
Once flashing, use the up and down keys to select
the desired run time. Press the "SET" key again to
adjust the frequency.
flashing, use the up and down keys to adjust the
frequency. NOTE: The frequency is only adjustable
in modes that include a shaking segment.
frequency permits the user to adjust the distance
travelled during the shaking segment. Choose a
small frequency for short distance (vibrating mixing).
Starting and Stopping the Rotator:
Once the settings are finalized, press the "Start/Stop"
key to begin the rotation. To stop the rotator before
the time has expired, press the "Start/Stop" key.
NOTE: The mixer may continue rotating for up to 1
second to return to the "home" position.
Once fully
V. Custom Modes (C1-C2)
The C1 & C2 modes can be used to create your own
personal mixing cycle by linking up to 6 mixing
functions together.
To begin, use the "MODE" key to select mode C1 or
(When
C2 and then press the "SET" key. Choose the first
mixing function by pressing the "MODE" key to
toggle
COUNTERCLOCKWISE rotating, PAUSE, FREQUENCY
or OFF.
COUNTERCLOCKWISE rotation, press the up and
down keys to determine the length of the rotation
(from 1 to 360º). When choosing a PAUSE function,
use the up and down keys to choose the length of
the pause in seconds (1 to 99 sec.) and when
choosing a FREQUENCY function, choose the length
of the frequency angle from 1 to 9. Once your first
mixing action has been configured, press the "SET"
key to advance to the next action.
setting, press "SET" to select the SHAKE TIME. The
shake time refers to the length of time the rotator
will conduct in FREQUENCY mode. Once selected,
press the "SET" key to advance to select the total run
TIME. Then press the "SET" key to store all selections.
Press the Up & DOWN keys to select the speed and
press the "START/STOP" key to begin the cycle.
When the frequency is
between
CLOCKWISE
rotating,
When choosing a CLOCKWISE or
After the sixth
The

Advertisement

loading

Summary of Contents for Benchmark R4040

  • Page 1 After the sixth After choosing the mode, the rotational speed can R4040: 29 x 13.7 x 19.9 cm (11.4 x 5.4 x 7.8 in.) setting, press “SET” to select the SHAKE TIME. The be selected by pressing the up and down keys to R4045: 41 x 13.7 x 19.9 cm (16.1 x 5.4 x 7.8 in.)
  • Page 2 Avoid the use of harsh solvents. manual configuration after each time that the stop key is pressed. Service and Contact: In the event that service or technical support is required, please contact Benchmark by phone at 1-908-769-5555 or by email at info@benchmarkscientific.com.

This manual is also suitable for:

R4045