Download Print this page

Sanyo MDF-U500VXC Product Features Manual page 6

-86°c ultra-low freezers
Hide thumbs Also See for MDF-U500VXC:

Advertisement

Twin Guard
Dual°Cool Refrigeration System
Dual, independent refrigeration systems operate in Normal or energy-saving EcoMode™ which is selected by the user and engaged through the
LCD controller. Normal Mode is used for the strictest of GMP applications. EcoMode™ is applicable to 90-95% of applications.
Twin Guard
Dual°Cool Refrigeration System
Status
System A and B ON
System A and B ON, Cycling On/Off
System A and B ON, Cycling On/Off
System A ON
System B ON
EcoMode
vs. Normal Mode
ECO mode operation image
Pull down
Stable temperature
Temp.
Comp A
Comp B
NORMAL mode operation image
Pull down
Stable temperature
Temp.
Comp A
Comp B
This chart shows the repeatable cycling wave form of normal mode vs. eco mode.
Normal mode is for strict GMP applications.
6
Twin Guard Series
-86°C Ultra-Low Freezers
Function
Maximum Pull-Down and Recovery Capcity
EcoMode™
Normal Mode
Back-Up for System B
Back-Up for System A
www.twinguardseries.com
Performance
Establishes highly uniform -86°C storage
temperature; maximizes recovery following
door openings and heat load additions
Maintains better energy management at
high or low ambient temperatures as well as
excellent top-to-bottom uniformity.
Maintains excellent top-to-bottom uniformity.
Maintains most repeatable, cycling wave form
for the strictest of GMP applications.
Maintains minimum -65°C reserve
temperature
Maintains minimum -65°C reserve
temperature
Chamber
temp
T12 on
T12 off
Comp A
Comp B
2 compressors are turned ON/OFF at same time
Chamber
temp
T on
T1 off
T12 on
Comp A
Comp B
Chamber
temp
T12 off
T12 on
T2 on
T2 off
Comp A
Comp B
2 compressors and fan motor for each are independently controlled by SV and read AT.
The SANYO Twin Guard Series
Freezer can be set for Normal or
EcoMode
operation. EcoMode
establishes an overlapping cycle to
significantly reduce energy consumption.
-86˚C
A
B
-86˚C
A
B
-86˚C
A
B
-65˚C
A
B
-65˚C
A
B
T2 off
T1 2 on

Advertisement

loading

This manual is also suitable for:

Mdf-u700vxc