Polydoros Lx; Polydoros Sx 65 - Siemens LITHOSTAR Multiline Installation Instructions Manual

Hide thumbs Also See for LITHOSTAR Multiline:
Table of Contents

Advertisement

2 - 2
Polydoros
SX 65
Polydoros
LX 30
AC
Fig. 1
Generator and Cabinet Installation
NOTICE
With the exception of the cooling unit and the high voltage charger, the cabinets are
shipped from the factory completely assembled and are individually crated.
We recommend setting up and bolting down the generator according to the project plan or
Figure 1 before installing the remaining cabinets.

Polydoros LX

Mark on the floor and drill the four mounting holes for the generator cabinet at its
intended location using the supplied template.
Drill the holes and insert the Liebig anchors that are included in the shipment (Fig. 2).
Remove all transport safety brackets from the generator.
Take the generator off the pallet.
If necessary, knock out the cable inlet in the generator cabinet for interconnections with
the adjacent cabinet. Install the edge protection strip (service package).
Mount the generator in place by inserting the 4 hex bolts into the Liebig screw anchors.
LITHOSTAR Multiline
System
N12
N11
M16
A13
A12
Do not insert the H1 high-voltage transformer until the generator
is cabled in the power cabinet.
RXL2-120.812.01
Rev. 02
System Installation
Cable lead-in
(on both sides)
Mounting holes
Cable lead-in from below
Fig. 2
Page 2 of 14
10.04
CS PS 24
Hand grip indentations
(on both sides)
2
Cable
lead-in from
the back
Siemens AG
Medical Solutions
2

Advertisement

Table of Contents
loading