R&S ZN-Z15 Series User Manual page 8

Calibration units

Advertisement

®
R&S
ZN-Z15x
The calibration unit relies on a USB voltage supply of 5V. If you connect
it via a USB hub, you have to use an active one.
During a firmware update of the network analyzer, the calibration unit
must be disconnected.
Do not disconnect the unit while data is being transferred between the
analyzer and the unit.
See also
"Cable selection and electromagnetic interference (EMI)"
on page 6.
3.5
RF connections
The calibration unit provides two RF connectors (numbered 1 to 2), to be connec-
ted to the test ports of the analyzer. The connector type is the same for all RF
ports of a calibration unit.
The maximum RF input power of the calibration unit is beyond the RF output
power range of the analyzer. If the device is directly connected to the test ports of
the analyzer, there is no risk of damaging the calibration unit.
See also
"Signal input and output levels"
User Manual 1337.5963.02 ─ 04
on page 7.
Preparing for use
RF connections
8