R&S ZN-Z15 Series User Manual page 11

Calibration units

Advertisement

®
R&S
ZN-Z15x
The calibration type depends on the number of ports and of the analyzer type. If a
single port is calibrated, the analyzer uses a reflection calibration type (e.g. Full
One Port / Refl OSM). For 2 and more ports, you can choose among several cali-
bration types.
4.4
User characterization
The calibration unit offers the possibility to store multiple user characterization
data files either on its internal flash memory or on the microSD card (R&S ZN-
Z152, R&S ZN-Z153). Use the "Characterize Cal Unit" function of your network
analyzer for this purpose. For details, refer to the analyzer's help system.
4.5
microSD card
On the rear of the R&S ZN-Z152 and R&S ZN-Z153 calibration units, there is a
microSD card slot. The microSD card can be used to store user characterization
data. The factory data is always stored on the internal memory.
Use the microSD card shipped with the calibration unit, if possible.
Before removing or inserting the microSD card, disconnect the USB
connection between calibration unit and network analyzer.
4.6
Mounting the calibration unit
To mount the calibration unit e.g. on a wafer prober, remove the elastic buffers on
the bottom of the housing. There are either two threads (R&S ZN-Z156) or four
threads for use of M2.5 screws.
User Manual 1337.5963.02 ─ 04
Working with R&S ZN-Z15x
Mounting the calibration unit
11