StarDental DentalEZ 263739 Instruction Manual

Autochuck turbine replacement kit

Advertisement

Quick Links

DO NOT USE FOR ORAL SURGERY
BEFORE INSTALLING THE TURBINE, CAREFULLY READ AND FOLLOW ALL OF
THESE INSTRUCTIONS. SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE
AND OBSERVE ALL WARNINGS AND CAUTIONS.
Turbine Installation
Note: Replace the turbine when the handpiece fails to achieve full speed and/or
increases sharply in noise level. The replacement cartridge is available from your
dealer and is easily replaced in the office. To order a replacement, see 'Replacement
Parts' section for StarDental part numbers.
1. Using wrench supplied, remove old end cap and turbine.
2. Inspect housing before cleaning. Remove any obvious debris to prevent damage
and to insure performance of the new turbine. Also, remove o-ring(s) from housing
at bur opening, if present.
3. Connect handpiece and run air through turbine housing without turbine installed
for 5 seconds.
4. Clean housing thoroughly with a cotton swab and isopropyl alcohol. Dry with syringe
air.
Clean debris from
inside of housing
5. Place new turbine assembly into housing and gently insert until fully seated.
Autochuck Turbine
Housing
6. Position and engage threads of end cap into the housing by hand until snug.
7. Tighten end cap with the combination end cap wrench, employing the wrench
opening marked AC as shown. Position wrench with marked side facing end cap
for best wrench control.
Wrench
Opening
Autochuck Operation
CAUTION
: Use burs that have an overall length of less than 24 mm and shank
length of at least 7.3 mm.
Autochuck Bur Changing System
1. Insert bur into autochuck until resistance is felt.
2. Depress autochuck end cap button and continue to feed-in bur until it clicks and
stops.
3. Release end cap button.
Now, apply light inward force to the dental bur. This will
help set the chucking mechanism for initial start up.
been installed to confirm that it is secured.)
4. To remove bur, grasp handpiece firmly so your thumb touches the end cap and push
until a slight 'click' is noticed.
5. Pull bur straight out.
6. Release endcap.
CAUTION
: Never depress the autochuck end cap button or rest the end cap
against any part of the oral anatomy during handpiece operation. Depressing the end
cap while a bur is rotating can activate the chuck mechanism and result in end cap
heat-up and/or loss of bur retention.
SAFETY
INSTRUCTIONS
1. Always tug on bur after chuck is loaded to confirm that the bur is secured.
2. Before use, verify concentric rotation by operating the handpiece outside the oral
cavity in a safe place. Do not use burs that do not run concentrically or produce
excessive noise.
3. Never operate handpiece without a bur installed.
4. Always use burs with an overall length of less than 24 mm. Shank length must be
at least 7.3 mm.
5. Use only recommended burs to prevent damage to autochuck and in some cases to
prevent burs from improper or incorrect placement. An improperly clamped bur may
dislodge during use, causing injury to you or your patient. A sudden, high-pitched
sound or increased vibration while handpiece is operating alerts the operator to the
possibility that the chuck is not clamped sufficiently to retain the bur or a bent bur
is inserted.
6. When using the autochuck, never depress end cap while bur is rotating.
7. For maximum life, never operate handpiece above 45 psi.
Always insert a bur the full length of the chuck. If you attempt to extend it from the
chuck, the bur may dislodge during use and cause injury to you or your patient.
Replace the turbine when the handpiece fails to achieve full speed and/or increases
sharply in noise level. This is an indication of a failing turbine. A failing turbine will reduce
the effiency of the handpiece and may result in end cap heat up. See "Turbine Installation".
Cleaning
(Always tug on bur after it has
Instructions
Note
After each patient, operate handpiece for 20 seconds with air
and water spray activated.
Cleaning
To remove debris from external surfaces, clean using a small
External
toothbrush with isopropyl alcohol. Dry thoroughly.
Surfaces
WARNING
CAUTION

Advertisement

loading
Need help?

Need help?

Do you have a question about the DentalEZ 263739 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for StarDental DentalEZ 263739

  • Page 1 To order a replacement, see 'Replacement Wrench is inserted. Parts' section for StarDental part numbers. Opening 6. When using the autochuck, never depress end cap while bur is rotating. 7. For maximum life, never operate handpiece above 45 psi.
  • Page 2 Lubricated Ceramic Bearing Autochuck Turbine 2 Years from date of insertion* 5. Place three (3) drops of DentaLube II, StarDental ® Part #262539, into drive air at rear Lubricated Steel Bearing Autochuck Turbine 1 Year from date of insertion* If any problem persists, contact dealer for further assistance;...

This manual is also suitable for:

Dentalez 264184Dentalez 265807Dentalez 263968Dentalez 263453Dentalez 260296Dentalez 260295 ... Show all