StarDental NuTorque Lite Operation Manual

Electric handpiece system

Advertisement

Quick Links

Please Read this operation manual carefully before use and keep for future reference.
CAUTION: Federal law restricts this device to sale by or on the order of a dentist.
Thank you for purchasing the NuTorque
maintenance guidelines. Keep this Operation Manual for future reference.
Application
This motor is used for cutting / polishing of the tooth.
User
Dentist and dental assistant.
Prohibition
This motor is dental treatment use only. Do not use for Oral Surgical treatment.
Classification of Devices
Type of protection against electric shock :
- Class II equipment :
Degree of protection against electric shock :
- Type B applied part :
Method of sterilization or disinfection recommended by the manufacture :
- Refer to 7. Sterilization
Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrousoxide :
- EQUIPMENT not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or
nitrous oxide.
Mode of operation :
- Continuous operation
Cautions for handling and operation
Read this manual thoroughly before use and operate the device properly.
These instructions will show you how to operate the product safety and prevent danger to you or others. They are
classified by the degree and/or severity of danger. All contents relating to safety should be oberserved.
Classification
W A R N I N G
Explains an instruction where personal injury or physical damage may occur.
C A U T I O N
Explains an instruction where minor to medium injury or physical damage may occur.
N O T I C E
Explains an instruction that should be observed for safety reasons.
Electric Handpiece System
OPERATION MANUAL
Lite. Read this Operation Manual carefully before use for operation instructions and care and
®
Degree and severity of danger or damage
OM-EZ505E

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NuTorque Lite and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for StarDental NuTorque Lite

  • Page 1 Electric Handpiece System OPERATION MANUAL OM-EZ505E Please Read this operation manual carefully before use and keep for future reference. CAUTION: Federal law restricts this device to sale by or on the order of a dentist. Thank you for purchasing the NuTorque Lite.
  • Page 2 3. · Connect to only 4-hole, 5-hole or 6-pin tubing. · NuTorque Lite system is to be used by qualified personnel for dental treatment only, and only in a dental office. · Patient safety must have first priority.
  • Page 3: Specification

    1. Specification AC Adaptor Control Box 265290(120V) and 265291(230V) Model M o d e l 265316 AC120V 50/60Hz 41VA Rated input AC28V 50/60 Hz Rated input AC230V 50/60Hz 41VA Drive Air Pressure 0.4MPa ( 4.0kgf/cm =56.89psi) or less Output AC28V 1.3A D i m e n s i o n s W127 x D149 x H66 mm AC120V FU100 : T1.6A 250V FU101 : T1.6A 250V...
  • Page 4 3. Operating Control Box CAUTION · Be sure that there is no air or water coming from Handpiece Tubing when attaching to control box. · Do not overtighten the handpiece tubing and motor cord nuts. ( 1 ) Connect the handpiece tubing from the delivery unit to the control box Handpiece Tubing Handpiece Tubing Fit the handpiece tubing connector from the delivery unit to the...
  • Page 5: Operation Section

    4. Handling the Motor CAUTION Ensure that the motor has completely stopped before handling the motor. ( 1 ) Connecting / disconnecting the motor and the motor cord Motor Cord Motor Cord Connector Connector To remove the motor cord from the motor, unscrew and detach the Motor Cord Motor Cord motor cord nut, and gently pull out the motor cord connector.
  • Page 6: Convenient Functions

    ( 4 ) Set the maximum speed using the speed control switches. ( Fig. 8) Set maximum Speed Set maximum Speed Speed Up Speed Up Speed Down Speed Down Fig.8 ( 5 ) Select FWD/REV rotation using the FWD/REV Select Key. (Fig. 9) Lamp illumniated only during REV rotation Lamp illuminated only during REV rotation Lamp ON...
  • Page 7 NOTICE " ( kgf/cm ) = 0.05MPa = 7.11psi. · Default value is set " · When the air pressure is not between " " - " ", " " is displayed, and the setting will not be memorized B. Setting Upper Limit: 1 ) After setting lower limit, press "...
  • Page 8 7. Sterilization Sterilize the Handpiece Attachments, Burs and Motor Only. Tighten For the sterilization method, we recommend the autoclave sterilization method. Note: Attachments, burs and motor Must Be Cleaned and Sterilized Before first use After Each Patient. Sterilizing the Motor 1 ) Turn off the Power.
  • Page 9 8. Maintenance Replace O-Rings if it becomes difficult to connect handpiece attachments or there is an air or water leak. ( 1 ) Changing the Oring Remove the O-Ring at the Motor Insert with a pointed tool, and mount the new O-Rings into the groove. CAUTION O-Ring ( Blue ) There are 4 O-Rings on the Motor.
  • Page 10: Error Code

    9. Error Code If the motor stops due to an abnormality such as a malfunction, overload, breakage or incorrect usage, the Control Box will display an error code on the Speed Indicator. If an error code is displayed, turn on the power again and check whether the same error code is displayed.
  • Page 11: Disposing Product

    Guidance and manufacturer's declaration - electromagnetic emissions The NuTorque Lite is intended for use in the electromagnetic environment specified below. The customer or the user of the NuTorque Lite and manual should assure that it is used in such an environment.
  • Page 12 TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the NuTorque Lite is used exceeds the applicable RF compliance level above, the NuTorque Lite should be observed to verity normal operation.
  • Page 13: Replacement Parts

    11. Replacement Parts Always insist on StarDental parts for best performance. Description Item Number NuTorque Lite Electric Handpiece System ..................265289 ® Control Box ............................265316 Motor ..............................265317 AC Adapter (120V) ..........................265290 AC Adapter (230V) ..........................265291 Motor Tubing ............................265326 O-Ring Kit for Motor ..........................

Table of Contents