Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. 1. Abnormal After Turn on The Infusion Pump 1.1 The alarm light on and indicating “use battery” after turn on the machine: Possible Cause 1: AC power cable is damaged or not plugged in.
Page 3
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. Diagram 3 Possible Cause 3: The driver board fuse was burnt. Solution: Use the multimeter to check if the fuse is connected; if no, use the same specification of fuse ( 2A 63V) for replacement. Location of fuse as shown in Diagram 4.
Page 4
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. 1.2 AC/DC indicator light, LCD screen and machine top indicator light are not on, and there is no boot tone after turn on the machine. Possible Cause 1: The connecting wire between main board and driver board is loosed or damaged.
Page 5
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. Possible Cause 2: The 5-pin connecting wire between power board and driver board is loosing or damaged. Solution: Please check the 5-pin connecting wire, reinserting or replacing it. Possible Cause 3: certain part of driver board is damaged.
Page 6
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. 1.3 AC/DC indicator light is on, but there is one or two abnormal between LCD screen, top indicator light, boot tone after turn on the machine. Possible Cause: Corresponding part of main board is damaged.
Page 7
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. Solution 2: Replace the connecting wire between main board and driver board and check if machine returns to normal. Possible Cause 2: driver board is damaged. Solution: Replace the driver board. 1.5Error code “0XE3” after turn on the machine.
Page 8
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. Diagram 11 Possible Cause 3) air bubble detection circuit on the motherboard damage Maintenance Step 1> contact the supplier . 2.2 Error code “0XE1” or “OXE2”during operation. OEX1 / OEX2 alarm means the wire connecting Hall sensor and pump body is loosing.
Page 9
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. Step 2: Insert the new Hall connecting wire...
Page 10
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. Step 3: Use solid sols to fix the connecting wire. The procedure to replace the Hall connecting wire is finished. 2.3 Always indicating low battery. Possible Cause 1: The battery is aging and can not recharge.
Page 12
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd. Diagram 14-2 Possible Cause The connecting wire between power board and driver board is poor contact or damaged. Solution 1: Open the front and rear shell, reinsert this connecting wire and check if machine returns to normal.