Testing The Transmitter - Philips 50 T Series Instructions For Use Manual

Fetal telemetry system
Hide thumbs Also See for 50 T Series:
Table of Contents

Advertisement

Testing the Transmitter

Testing the Transmitter
56
1. Slide back the battery cover.
2. Switch on the transmitter. The green On/Off light (A) comes on
showing the transmitter is on.
3. Check the red light (B) situated behind the middle battery. If:
• the red light is lit for a few seconds and then goes out, the self
test is successfully complete.
• the red light blinks, or remains on after three seconds, change
the batteries. If the fault continues, contact your Philips
Medical Response Center.