Subject Breath Test Procedure

Step by Step
STEP 1 – Preliminary questions to the subject
As far as possible, ensure the subject has taken nothing by mouth
(alcoholic or otherwise) in the last twenty minutes, and has not smoked in
the last two minutes. If necessary, wait.
STEP 2 – Switch on
Press Confirm / Action Button (C) and hold for approximately 2 seconds
until the operator hears a beep. The instrument will scroll through its
start-up screens displaying the instrument temperature, time, date, serial
number and the software version of the instrument. If the instrument is
within 15 days of the service lock, the start-up screens will be followed by
a warning message showing the next service date and the number of days
remaining.
STEP 3 – Function selection
Within the OPERATOR MENU there are two breath test options, a last test
recall option, an option to power down the instrument and an option to
access the Supervisor menu.