Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 2 of 46 Page Document title HygroGen 2 - Quick Start Guide: Congratulations on selecting the HygroGen humidity and temperature generator; the unit will enable you to perform accurate and rapid humidity and temperature calibrations for many years.
HygroGen Settings Section. The convenience of PC functionality and a USB interface enables the use of Rotronic HW4 software and USB peripherals such as a keyboard and mouse. Nine USB connections are fitted to the front and back of the HygroGen.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 5 of 46 Page Document title Part and Screen Identification Diagrams Figure 1: Front of HygroGen Power switch USB ports Touch screen controller...
Page 6
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type Page 6 of 46 Document title Figure 3: HygroGen Screens The HygroGen controller consists of four primary screens that are accessed using the forward or back buttons in a loop: Secondary Screens are accessed through touch screen buttons.
Page 7
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 7 of 46 Page Document title Figure 4: Controller Primary Screen Menu access (top left corner HOME) Parameter identification °C Actual chamber temperature...
Page 8
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type Page 8 of 46 Document title Figure 7: Program Edit Screen Graphical display of program Program step number Temperature set-point Humidity set-point...
‘How to Fill the Water Reservoir’ - Section 2.1 (page 13) If the unit is used in a manner not specified by ROTRONIC, the function and protection provided by the unit may be impaired.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 10 of 46 Page Document title 1.4 Setting Temperature and Humidity Set-Points The temperature and humidity set-points are set by touching the set-point value on the primary controller screen.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 11 of 46 Page Document title Humidity Map 1.6 Above Ambient Temperature Operation At high temperature and high relative humidity conditions it is necessary to allow the unit to stabilize to ensure minimal condensation.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 12 of 46 Page Document title Play/stop – Starts or stops the program and displays the Run and Step timing information Program Run Time –...
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 13 of 46 Page Document title 1.9 Low Water Level Alarm The HygroGen requires water to create high levels of humidity returning from the internal saturator.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 14 of 46 Page Document title Section 2: Maintenance and Servicing 2.1 How to Fill the Water Reservoir The HygroGen requires 30…80 ml of distilled water in the internal reservoir. Insufficient water will deactivate the humidification system.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 15 of 46 Page Document title 2.3 How to Empty the Water Reservoir 1) Place the HygroGen on a level surface.
Dusty desiccants should be avoided as the dust may be passed in to the HygroGen chamber. Material Safety sheets for the desiccants supplied by ROTRONIC are available on request. 2.6 Molecular Sieve Desiccant (Factory Standard) Molecular sieves have the appearance of small opaque tan beads.
For most situations it recommended that new silica gel is used unless a proven process for regeneration can be implemented. Regeneration can be carried out repeatedly, although eventually the beads will lose their colour. Replacement Orange to Green self-indicating silica gel can be ordered from ROTRONIC. 2.10 Drierite Desiccant Drierite is anhydrous calcium sulphate.
The structure of the Drierite granules can break down during regeneration. If the result is dusty it should not be reused in the desiccant cell. Replacement Drierite desiccant can be ordered from ROTRONIC. 2.12 Blue-to-Pink Self-Indicating Desiccant Until recently most self-indicating desiccants used a small concentration (between 0.5 and 3.0 % weight...
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 19 of 46 Page Document title 2.15 HygroGen External Cleaning Prior to cleaning, the unit must be disconnected from the electrical mains supply, as described in Power Isolation Section 4.2.
(Section 13). For optimum performance ROTRONIC recommend that the HygoGen is serviced on an annual basis (every 12 months) through an approved HygroGen service centre, details of which can be found in Section 13 of this manual.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 21 of 46 Page Document title 2.18 Water Quality For best performance, HygroGen should only be filled with distilled water. As a precaution against any microbial and algae water contamination, HygroGen incorporates an Ultra Violet (UV) sterilisation lamp within the water reservoir.
‘HW4 embedded’ for information on the functions available. 3.3 Rotronic USB Devices With HW4 running on the external monitor, it is possible to connect any Rotronic product with USB interface to HW4 using any of the HygroGen’s USB ports. Simply connect the device and use the HW4 ‘Scan for USB Masters’...
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 23 of 46 Page Document title 3.4 Connecting a Chilled Mirror Instrument Chilled mirror instruments offer the best possible reference measurement thanks to their inherent long term stability.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 24 of 46 Page Document title Section 4: Installation, Safety and Transit 4.1 Installation To ensure correct operation please ensure that: 1) The unit is level and stable;...
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 25 of 46 Page Document title 4.8 Preparing the Unit for Transit or Storage When moving the HygroGen around site between calibrations, it is not necessary to empty the water reservoir (Section 2.3) provided the unit is maintained in a horizontal orientation and handled carefully.
Incorrect power down or unplugging the HygroGen while in use may cause software corruption. This may mean the HygroGen2 will not start. In this event, contact your supplier or nearest HygroGen service centre; they will be able to provide you with a bootable USB disc that can be used to restore the HygroGen to the factory condition.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 27 of 46 Page Document title 4.10 Shut Down Procedures There are two shut down procedures; these are presented after you press the on screen power button.
Section 3.2). Any Rotronic USB probes that are connected to the HygroGen’s USB ports can be connected to HW4 using the ‘Search for USB Masters’ command. For a full overview of the use and features of HW4, please refer to the full HW4 manual.
The following Sections and the uncertainty budgets included are for guidance only, and ROTRONIC accepts no liability whatsoever in relation to their use. Almost certainly, your need to calibrate is based on the need to meet the requirement of traceable calibration within a quality system.
Document title 6.2 Internal RH Probe Reference HygroGen uses a ROTRONIC HygroClip RH and temperature probe for control. The HygroClip provides a digital signal to the HygroGen control system, so there can be no degradation of the measured value due to (for example) analogue to digital conversion. HygroGen’s displayed measurement values can...
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 31 of 46 Page Document title 6.3 External RH Probe Reference An external RH reference is applied through the HygroGen door in the same way as an instrument under test.
Document title 6.4 Chilled Mirror Humidity Reference For the lowest possible calibration uncertainty, the ROTRONIC recommendation is to use a chilled mirror instrument to provide the best possible reference humidity (contact ROTRONIC for details of tested chilled mirror systems). This type of instrument is generally accepted as a secondary humidity standard.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 33 of 46 Page Document title Section 7: Practical User Notes 7.1 Probe Insertion Depth When adjusting or calibrating one or several probes against a reference probe (such as a PRT), be sure to immerse all probes (including the reference probe) to the same depth.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 34 of 46 Page Document title 7.4 Desiccant Choice Molecular sieves should be preferred to other desiccants. Molecular sieves can be fully re-generated, time and time again, simply by exposure to 150 ˚C for 4 to 5 hours (after removal from the desiccant...
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 35 of 46 Page Document title Section 8: HygroGen Settings and Updates 8.1 HygroGen Settings As well as using the forward >> and backward << keys, you can access each menu from a drop down menu in the top left-hand corner.
Rotronic service centre. The upgrade may be downloaded from the Rotronic website, or posted. In both cases it should be saved on a USB memory stick that does not contain any additional files, especially with .exe or .ini extensions. If in any doubt, refer to a Rotronic service centre for support.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 37 of 46 Page Document title Section 9: HygroGen Screen Messages On Screen Message: Fault/Action: HygroClip NC HygroClip not connected or faulty, check and replace as required. Contact service centre if message persists.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 38 of 46 Page Document title Section 10: HygroGen Specification HygroGen 2 Specification Relative Humidity Temperature Control Sensor HygroClip2-S, capacitive RH sensor, Pt100 temperature sensor.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 39 of 46 Page Document title Section 11 HygroGen Order Codes: Order code: Description HG2-S HygroGen with touch screen user interface, set-point control and programmer function, heated sample loop, 1x desiccant cell, 1x fill syringe, embedded HW4-P software.
Page 40
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 40 of 46 Page Document title xx probe diameter codes not used 15mm (HygroClip) 13.5mm 20mm 26mm 25mm 34mm 10mm 14mm...
Section 12: HygroGen Warranty Statement ROTRONIC will grant a warranty for the HygroGen for a period of 12 months from the date of delivery in respect of any evidenced faulty workmanship and materials. The warranty period is extended by a further 12 months when the unit is serviced at a Rotronic approved service centre 12 months after the date of supply.
Rotronic AG E-M-HG2-S-V1_04 Bassersdorf, Switzerland Document code Unit Instruction Manual HygroGen2 humidity and temperature generator: Instruction Manual Document Type 42 of 46 Page Document title Section 13: Manufacturer & Service Centre Contact Information Manufacturer/UK Service Switzerland Service USA Service Centre:...
Page 43
1.59 Drift between two successive calibrations of the reference instrument Resolution from manufacturers specification Measured temperature gradients from the Rotronic HygroGen Service Measured stability from the Rotronic HygroGen Service From Standard Deviation of 10 repeated calibration runs Resolution of Rotronic HygroGen Control Clip...
Page 44
Expanded uncertainty (95 % confidence) 1.69 Drift between two successive calibrations of the reference instrument Resolution from manufacturers specification Measured temperature gradients from the Rotronic HygroGen Service Stem conduction Measured stability from the Rotronic HygroGen Service From Standard Deviation of 10 repeated calibration runs...
Page 45
From Standard Deviation of 10 repeated calibration runs Expanded uncertainty (95 % confidence) 0.77 Measured temperature gradients from the Rotronic HygroGen Service Measured stability from the Rotronic HygroGen Service From Standard Deviation of 10 repeated calibration runs Resolution of Rotronic HygroGen Control Clip...