Operating PC LM Control Unit
Fig. 5
Open operating PC and install board D51
•
Remove all external cables.
•
Remove the upper control unit cover.
•
Remove the Slot 1 cover plate.
•
Install and fasten board D51 in Slot 1 (1/Fig. 5).
•
Plug reset cable on board D51.X3 Pin 4 (1/Fig. 6) and on the motherboard at CN4 Pin 1
(2/Fig. 6).
•
Reconnect all cables except for DUEP X51 to the PC.
•
System ON.
•
Check that the display on the board D51 runs back from "9" to "1".
- The display remains for a short time at "3"
- "0" = everything OK
- "1" = no DUEP comes
•
System OFF.
•
Connect the DUEP cable X51.
Close the operating PC
•
Fasten the upper control unit cover again.
•
Fasten the control unit to the drawer.
•
Push in the drawer paying attention to the cable run in the console.