Check whether the display on board D51 runs backwards from "9" to "1".
- The display stops temporarily at "3"
- "0" = OK
- "1" = no DUEP
Switch the system OFF.
Connect the DUEP cable X51.

Closing the operating PC

For the upper part of the housing:
- place it on the operating PC in the direction of the arrow (1/Fig. 7).
Ensure that the anchor pins on the inner side of the upper part of the housing lock into
the corresponding guides.
- press in the direction of the arrow (2/Fig. 7), until it noticeably locks into place.
Push in the drawer; be careful of the cable layout in the operating console.
Push the operating PC to the back and right in the operating console, so that the ventila-
tion ducts are not blocked.
LITHOSTAR Multiline
System Manual
Operating PC SCENIC Pro C5/ Pro C6
Register 5
RXL2-120.091.08
Rev. 04