Fisherbrand Accu100 Instruction Manual

Fisherbrand Accu100 Instruction Manual

Ultrapure water system
Hide thumbs Also See for Accu100:

Advertisement

Quick Links

Accu100
Ultrapure Water System
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Accu100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Fisherbrand Accu100

  • Page 1 Accu100 Ultrapure Water System USER MANUAL...
  • Page 2 Notice The Accu100 system is manufactured in a facility whose quality management system is approved by an accredited registering body to the ISO9001 Quality System Standards. DECLARATION OF CONFORMITY EUROPEAN UNION EC DIRECTIVES AND FCC RULES Directive 2006/95/EC FCC Part 15 Class B...
  • Page 3 This Manual is Applicable to the Following Accu100 Models 15489749 Accu100 Ultrapure Water System, 100 - 240 VAC 15499749 Accu100 UV Ultrapure Water System, 100 - 240 VAC www.eu.fishersci.com/fisherbrand...
  • Page 4: Table Of Contents

    4.2 Maintenance Menu and Settings ..............25 4.3 Replacing the Ultrapure Pack and the Final Filter ........27 4.4 Replacing the UV Lamp (for UV models) ........... 29 5 BASIC TROUBLESHOOTING ................32 6 PARTS AND ORDER INFORMATION .............. 33 7 WARRANTY INFORMATION ................35 www.eu.fishersci.com/fisherbrand...
  • Page 5: Introduction

    1) Use with a properly grounded electrical outlet of correct voltage and current handling capacity. 2) Replace fuses with those of the same type and rating. 3) Disconnect from the power supply prior to maintenance and service. 4) Refer servicing to qualified personnel. www.eu.fishersci.com/fisherbrand...
  • Page 6: Product Features

    The Accu100 Ultrapure Water System produces ultrapure water from reverse osmosis (RO), distillation or deionized water. Ultrapure water from the Accu100 system meets or exceeds quality standards of ASTM, CAP, CLSI and ISO Type I water standards. The system has the following characteristics: •...
  • Page 7: Main Applications

    Particles in Ultrapure Water < 1 /mL (with a 0.2 µm final filter or final (> 0.2 µm) ultrafilter) < 0.1 cfu/mL (with a 0.2 µm final filter or final Microorganism ultrafilter) Pyrogen Content < 0.001 EU/mL (with a final ultrafilter) www.eu.fishersci.com/fisherbrand...
  • Page 8: Installation

    Scissors or a box opener to open packages and cut water tubing 2.2 Items Included The Accu100 system comes with the following items in the package. The following items are packed separately. These items need to be installed at initial set-up:...
  • Page 9: Structure And Flow Diagrams

    2.3 Structure and Flow Diagrams 2.3.1 System External Layout Control panel Display panel (see explanation on 2.3.3) Lock bar Final filter Purification cartridge Front View Lock bar Power switch Final filter Power Socket Handle Purification cartridge Water inlet Side View www.eu.fishersci.com/fisherbrand...
  • Page 10 2.3.2 Accu100 System Water Flow Diagram The water flow path for the Accu100 and Accu100 UV models are described below. Each component is explained in the legend. ACCU100 UV Flow Diagram (DISTILLED WATER) (RO WATER) (DI WATER) I T E M...
  • Page 11 In Ready mode, press it to enter Recirculation Recirculation mode, Press simultaneously to enter user menu. In Ready mode, press it to enter volumetric Volumetric dispensing. Press simultaneously Dispensing to enter user menu. Pressure In Ready mode, press it to enter Pressure Relief Relief mode. www.eu.fishersci.com/fisherbrand...
  • Page 12: Installation

    (push in) Please follow the instruction below to remove the stopper A. Insert the tweezers to the gap between the stopper and the collet. B. Pinch the tweezers and push it as the picture indicated to remove the stopper. www.eu.fishersci.com/fisherbrand...
  • Page 13 Blue caps 2.4.1.2 Open the lock bar at the front of the system. 2.4.1.3 Remove the two blue caps at the water inlet/outlet inside the cabinet. Stoppers 2.4.1.4 Wet the O-rings on the cartridge with purified water. O Rings www.eu.fishersci.com/fisherbrand...
  • Page 14 1) Insert one end of the PE tubing into the water inlet (Note: make sure to insert the tubing all the way in for about 20 mm). 2) Gently pull the tubing outwards to check if Water inlet the connection is secure. www.eu.fishersci.com/fisherbrand...
  • Page 15 1) Make sure the power switch is OFF at "O" Power cord position. 2) Insert one end of the power cord into the socket on the left side of the Accu100 system. 3) Plug the other end of the power cord into the power outlet. Note: Do not install the final filter at this stage.
  • Page 16: System Start Up And Operation

    Prior to system start up, use the table below to make sure all parts have been installed and connected, and quality of feed water meets minimal requirements. Check List Results Feed Water quality Water temperature: 5 - 35 Feed Water connection Power cord connection AccuQuatro U Pack cartridge installed www.eu.fishersci.com/fisherbrand...
  • Page 17: System Start Up And Display

    Press to dispense ultrapure water, system enters Product mode. Product In Product mode Press again to stop dispensing. 18.2 MΩ.cm System enters Recirculation mode. In Recirculation mode Recirculation Press to dispense water. 18.2 MΩ.cm Press to enter Ready mode www.eu.fishersci.com/fisherbrand...
  • Page 18 Screw the degassing fitting from the system accessory bag onto the water outlet and connect it with 1/4 inch PE tubing, put another end of the tubing to the sink. Press to enter Product mode; Gas inside the system will be flushed out with water. www.eu.fishersci.com/fisherbrand...
  • Page 19 Knob for discharging trapped air 3.2.4.2 Screw the final filter onto the water outlet till finger tight (no leaking at dispensing). DO NOT over-tighten it. 3.2.4.3 Loosen the knob on the top of the final filter for degassing. www.eu.fishersci.com/fisherbrand...
  • Page 20 Recirculation into Ready mode. 2) Installation is completed. 3) Alternatively, the system automatically enters Ready mode after 30 s without pressing If an ultrafiltration filter for pyrogen removal is needed, follow the same installation procedures as above. www.eu.fishersci.com/fisherbrand...
  • Page 21 UV and press to reset UV lamp FLOW: 1.50 L/min UNIT: MΩ.cm life. Newly replaced UV lamp shows 720 TIME: 2014-11-27 09:21:27 LANGUAGE/语言:ENGLISH days life. SERIAL NO: S4PPAL3701 Press simultaneously for 3 seconds again to exit user menu. www.eu.fishersci.com/fisherbrand...
  • Page 22: Routine Operation

    3.3 Routine Operation Turn on the power switch. Press to put the system from Ready to Recirculation mode and wait for the resistivity to reach 18.2 MΩ·cm. Press to dispense water. Press again to stop dispensing, system enters Recirculation mode. www.eu.fishersci.com/fisherbrand...
  • Page 23: Volumetric Dispensing

    3.4 Volumetric Dispensing 3.4.1 In Ready mode, press to enter volumetric dispensing mode. Volumetric 01.0 L 3.4.2 Press to adjust the volume. Volumetric 02.0 L 3.4.3 Press to start dispensing. 3.4.4 System enters Recirculation mode after dispensing. Recirculation 18.2 MΩ.cm www.eu.fishersci.com/fisherbrand...
  • Page 24: Maintenance

    UV lamp has reached the end of its expected EXCH. UV LAMP Yellow service life and need to be replaced. Accu100 system can be connected to a tank level sensor. When water level inside the tank TANK EMPTY is below the minimum level, system will not...
  • Page 25: Maintenance Menu And Settings

    UNIT: MΩ.cm TIME: 2014-11-27 09:21:27 LANGUAGE/语言:ENGLISH SERIAL NO: S4PPAL3701 4.2.3 Re-setting UV Lamp PACK: 360 DAYS Please follow the procedures in 3.2.5.3 FILTER: 360 DAYS UV: 720DAYS FLOW: 1.50 L/min UNIT: MΩ.cm TIME: 2014-11-27 09:21:27 LANGUAGE/语言:ENGLISH SERIAL NO: S4PPAL3701 www.eu.fishersci.com/fisherbrand...
  • Page 26 TIME. UV: 720DAYS FLOW: 1.50 L/min UNIT: MΩ.cm TIME: 2014-11-27 09:21:27 Press to move cursor to LANGUAGE/语言:ENGLISH corresponding number and press SERIAL NO: S4PPAL3701 increase the numbers. Press simultaneously to exit and back to Ready mode. www.eu.fishersci.com/fisherbrand...
  • Page 27: Replacing The Ultrapure Pack And The Final Filter

    Press to depressurize the system. d. When no water is coming out from the dispense port, press to put the system into Ready mode. Screw off the degassing fitting. www.eu.fishersci.com/fisherbrand...
  • Page 28 Follow procedures in 3.2.3 for degassing. h. Press to put system into Ready mode. Follow procedures in 3.2.5.1 to reset AccuQuatro U Pack life. Follow procedures in 3.2.4 for installing final filter. k. Follow procedures in 3.2.5.2 to reset final filter life. www.eu.fishersci.com/fisherbrand...
  • Page 29: Replacing The Uv Lamp (For Uv Models)

    UV lamp by hands. Gloves are included in the UV lamp package. a. Confirm the system is in Ready mode. Otherwise, press to put the system into Ready mode. b. 1) Turn off the power. 2) Unplug the power cord. www.eu.fishersci.com/fisherbrand...
  • Page 30 2) Unplug the UV lamp from its power cord. 1) Hold the top of the new UV lamp and connect it to the ballast cable connector (4-pin connector), 2) Gently insert the UV lamp completely into the chamber. www.eu.fishersci.com/fisherbrand...
  • Page 31 Recap the UV lamp chamber with the black cover on the power cable. h. Reinstall the system back cover and tighten the screws. Do not over-tighten them. Follow the procedures in 3.2.5.3 for resetting UV lamp. www.eu.fishersci.com/fisherbrand...
  • Page 32: Basic Troubleshooting

    Contact a ULS agent Low water flow rate Final filter clogged Replace final filter System internal water Contact a ULS agent flow clogged If the problem occurring is not listed in the chart above, please contact a ULS agent. www.eu.fishersci.com/fisherbrand...
  • Page 33: Parts And Order Information

    Description 15489749 Accu100 Ultrapure Water System, English, 100 - 240 V 15499749 Accu100 UV Ultrapure Water System, English, 100 - 240 V 15479709 AccuQuatro U Pack Cartridge (RO Feed) AccuQuatro U Pack Cartridge kit with a 0.2 µm Capsule 15489709...
  • Page 34 Other Maintenance Spare Parts ontact Fisher Scientific for ordering information: Main control panel Booster Pump Resistivity Sensor Commonly Used Connector Package UV lamp ballast Power Switch Fuse 3/8" PE tubing (5 meters) 1/4” PE tubing (5 meters) www.eu.fishersci.com/fisherbrand...
  • Page 35: Warranty Information

    Thermo Fisher Scientific assumes no responsibility for any errors that may appear in this manual. The plant's quality management system where this Accu100 water system was manufactured has passed the ISO9001:2008 quality management system certification.

Table of Contents