Contents Introduction .....................4 Important information ..................5 System information..................... 5 Application and classification ................5 Liability ....................... 5 Scope of delivery....................6 Safety instructions ..................7 Description of the device................8 Composition of the slit lamp ................8 Microscope ......................9 Slit projector ..................... 10 Setting-up, assembly and repair ..............11 Powering-up ....................13 Operation of the apparatus ................14...
Note! Indicates information for correct use so that errors may be avoided. This publication may not be copied or transferred without prior agreement from bon Optic. bon Optic reserves the right to make changes in the interest of technical development. These operating instructions are not subject to updating.
: bon SL-72/SL-75 2.2 Application and classification The bon SL-72/SL-75 slit lamp is used to examine the structures at the front of the eye. The SL-72/SL-75 slit lamp is a Class 1 non-invasive, active medical device in accordance with the classification regulations of Directive 93/42/EWG on medical devices (MDD).
2 Important information bon SL-72 / SL-75 2.4 Scope of delivery 1 x slit lamp bon SL-72/SL-75 1 x instruction manual 1 x dust shield 1 x focussing light 1 x replacement lamp 6V/20W 1 x non-slip surface 1 x mouth cover with fixing screw...
3 Safety instructions bon SL-72 / SL-75 3 Safety instructions Please follow the legal requirements on accident prevention and observe the following safety instructions! Setting-up and assembly: The slit lamp should not be assembled and operated in damp rooms. Ensure that the device is on a level and stable surface during assembly.
4 Description of the device bon SL-72 / SL-75 4 Description of the device The bon slit lamp SL-72/SL-75 is a professional universal instrument with magnification adjustment for examination of the front structures of the eye. 4.1 Composition of the slit lamp...
(on SL-75) (3) Eyepiece (replaceable) Diagram 4-2: Microscope of the SL-72/SL-75 The eyepiece (3) with standard 12.5x magnification enables an ametropia equalisation of +/- 5 dpt. By extending or pushing in the eyepiece lever (4), the distance to the pupil can be adjusted.
5 Setting-up, assembly and repair bon SL-72 / SL-75 5 Setting-up, assembly and repair a) Screw the gearwheel plates onto the table base with two recessed-head screws each. Ensure that the sides of the gear wheels face outwards! b) Screw the Teflon non-slip surface to the table base with four recessed-head screws.
6 Powering-up bon SL-72 / SL-75 6 Powering-up The slit lamp can be supplied with electricity by a separate mains adaptor or via an examination unit. a) Original mains adaptor: (1) Mains connection and fuse (2) Focussing light connection (3) Slit lamp connection...
7 Operation of the apparatus bon SL-72 / SL-75 7 Operation of the apparatus 7.1 Focussing the microscope Remove the cap and place the test rod in the holder as shown. Turn the slit lamp on. Focus the eyepiece so that the grainy surface of the test rod can be seen in focus at medium magnification.
Do not use disinfection alcohol! 8.2 Maintenance When used properly, the slit lamp should not require repeated maintenance. For repairs or technical problems please contact the bon Optic customer services department. 15/21...
8 Maintenance and care bon SL-72 / SL-75 8.3 Repairs you can carry out yourself 8.3.1 Changing the slit projector lamps Important The type of lamp used can vary depending on the version of the product. Please check which type of lamp is...
3) Remove the old fuse from the holder and replace it with a new fuse of the same type. 4) Place the fuse holder back in the fuse socket. 5) Close the fuse element. 8.4 Power supply disruptions Should EMC disruptions (electro-magnetic compatibility) occur from the mains adaptor, please contact bon Optic customer support. 17/21...
A receipt with the date of purchase can be provided. The device has been used properly and in accordance with the conditions of use. Repairs have not been carried out by anyone other than the bon Optic customer service team or persons authorised by bon Optic.
This symbol, found on your product or on its packaging, indicates that this product should not be treated as household waste when you wish to dispose of it. Instead, it should be handed over to an applicable collection point for the recycling of electrical and electronic equipment. If you have any further questions, please contact bon optic.
Page 20
10 Technical data bon SL-72 / SL-75 Diagram 10-1: Slit lamp components Focussing light 21) Scale for slit projector Eyepiece lever 22) Screw for moveable arm Eyepiece (replaceable) 23) Holder for test rod Screw for mouth cover 24) Screw-top lamp housing...
Page 21
Gerätetyp / UMDNS-CODE: Spaltlampe (12-281) (Device type/ UMDNS-CODE) Slit lamp (12-281) Spaltlampe SL-72/SL-75/SL-85/SL-E/SL-Zoom Gerätebezeichnung: Slit lamp SL-72/SL-75/SL-85/SL-E/SL-Zoom (Device name) Klassifizierung: 1 (Richtlinie 93/42/EWG, Anhang IX, Regel 1) (Classification) 1 (MDD 93/42/EEC, appendix IX, rule 1) Wir erklären hiermit die Übereinstimmung des vorgenannten Produkts mit der EU-Richtlinie 93/42/EWG über Medizinprodukte.