TaKaRa PCR Thermal Cycler Dice
Home Scree
PCR Program Editor:
Configure the PCR Program
PCR File Manager:
For managing the PCR run files on the device
Manual Incubation:
Use for incubator
 
Configuration:
Optional settings of the unit
The Power Switch locates on the rear of the unit to put on or shut off the Power supply.
※ Avoid to touch the screen and shut off the power supply until the screen appear the 'Home' screen. That is the cause of failure the
  unit and wrong movement.
1.Prepare new user acountment and folder
2.Create new PCR program
3.Edit the PCR program
4.Load the configured PCR program
5.Setting of specific measurement
6.Manual Incubation
.Optional setting
【Caution】・ The beginning to use TaKaRa Thermal Cycler DiceTM Touch, you may read this Quick Guide and User's
      ・ Instructions and installations about TaKaRa Thermal Cycler Dice
     ・ The stylus[touch pen] (that is optional of the unit) is used to operate the program on the Touch Panel.
TM
Touch  
・・・
TM
Touch is confirmed the content of User's Manual .
TAKARA BIO INC.
Ver.2.0.1
・・・2
・・・
・・・
・・・
・・・
・・・
・・・

Advertisement

loading

Summary of Contents for Takara Bio Thermal Cycler Dice Touch

  • Page 1 TAKARA BIO INC. TaKaRa PCR Thermal Cycler Dice Touch   Ver.2.0.1 Home Scree PCR Program Editor: Configure the PCR Program PCR File Manager: For managing the PCR run files on the device Manual Incubation: Use for incubator   Configuration: Optional settings of the unit The Power Switch locates on the rear of the unit to put on or shut off the Power supply.
  • Page 2 Dice Touch Quick Guide -2 TAKARA BIO INC. 1.Prepare new user acountment and folder Prepare new user accountment Touch a 'Configuration' from the Home screen.       Touch an 'User Accounts' icon from Configuration. Touch an 'Add User' on the operating icons under the user accounts screen.
  • Page 3 Dice Touch Quick Guide -3 TAKARA BIO INC. 2.Create new PCR Program Touch the 'PCR Program Editor' from the Home screen. Show a standard program(named Program1) on the screen. Touch the 'Wizard' icon of PCR Program Editor. Edit the 'PCR Experiment Wizard' screen.
  • Page 4 Dice Touch Quick Guide -4 TAKARA BIO INC. 3.Edit the PCR program Touch a 'PCR Program Editor' from Home screen. Change the User Accounts to click this part. Show a standard program on the screen. Touch an 'Load File' icon on the bottom shelf.
  • Page 5 Dice Touch Quick Guide -5 TAKARA BIO INC. Check 'Add to end of Profile' or 'Insert at start of Profile' or Insert after xx without cycling. The first screen of the Add Event Wizard: ● Cycling (2~10 Step) ● Hold ● Ramp (≦2.5℃/sec) ● Pause...
  • Page 6 Dice Touch Quick Guide -6 TAKARA BIO INC. Touch a button of 'Preheat' to preheat top heater. Press OK✔ to turn top heater ON, CANCEL× to turn top heater OFF on System Message. Change the 'Preheat' to red button of 'Heating' with top heater heating.
  • Page 7 Dice Touch Quick Guide -7 TAKARA BIO INC. 4.Load the configured PCR program Touch a 'PCR File Manager' from Home screen. Select an User Account for use or an USB drive from Accounts on the PCR File Manager. Appear some program files(.t3p) at the User Account for use.
  • Page 8 Dice Touch Quick Guide -8 TAKARA BIO INC. Copy: Copy the program files. Copy plural program files surrounding them with a drag. Paste: Paste the copied program files. Select an account that want to make copied files. Delete: Delete the program files.
  • Page 9 Dice Touch Quick Guide -9 TAKARA BIO INC. 5.Setting of specific measurement Touch a Step Options/Advanced Edit. 1.Setting of Gradient Set different temperatures hat is programmable 0~25℃ across block width (12wells) for both the left and right sides of the block.
  • Page 10 Dice Touch Quick Guide -10 TAKARA BIO INC. 2.Setting of Touchdown The means of prevent non-specific amplification from PCR reaction. Check the 'Touchdown' checkbox. Given step within a cycling event to be progressively decreased or (increased) from an initial value to a final value over a defined number of steps.
  • Page 11 Dice Touch Quick Guide -11 TAKARA BIO INC. 6.Manual Incubation     Touch a 'Manual Incubation' from Home screen. Touch a column of Temp, Enter target incubate temperature then click run. Top heat : Check a checkbox of Top heat on the Run control.
  • Page 12: Optional Setting

    Dice Touch Quick Guide -12 TAKARA BIO INC. 7.Optional setting Touch a 'Configuration' from Home screen. 1. System Information ① ③ ② ④ 2. Machine Settings 3. Time & Date 4. Thermal Calibration 5. User Accounts 6. Software Upgrade ⑤ ⑥ ⑦ 7. Touchpanel Calibration 1. System Information Confirm the situation of the unit.
  • Page 13 Dice Touch Quick Guide -13 TAKARA BIO INC. Used to adjust the machine calibration. 4. Thermal Calibration This machine is factory calibrated and should not require user recalibration of the machine. Click a 'Ramp Rates' button on the below column of the Calibration Wizard. Appear the ramp rates of the temperature curve just before the reaction.
  • Page 14 TAKARA BIO INC. List of supplies( Tubes and Plates) 0.2 ml Single Tube 0.2ml Hi-Tube Dome Cap Recovery ※ 0.2ml Hi-Tube Flat Cap Recovery ※ 0.2 ml 8-Tube&8-Cap 0.2ml Hi-8-Tube (8-Tube) 0.2ml Hi-8-Flat Cap (8-Cap) 0.2 ml Plate & Cap 96 well snap plate (12×8, 96well Plate)...