The instrument can be applied for the extraction and purification of nucleic acid from different clinic samples. Purifier HTS is especially used on research and tests by the personnel who have been trained professionally. It is used for automatic transfer and processing of magnetic beads in deep well plates.
Different from the traditional nucleic acid extractor, Purifier HTS contains two heating plates and two standard plates. Only four plates can complete the complete steps of nucleic acid extraction by magnetic bead method, so as to realize the high-quality extraction of nucleic acid in the sample, which greatly shortens the extraction time.
• Humidity is low and condensation does not occur (relative humidity is between 10% and 80%); • Install the Purifier HTS in a protected place where no one can step on or trip over the power cord, and where it is easy to access the power cord when the plug needs to be removed;...
Chapter 4 Instructions for Use 4.1 Starting Before turning on Purifier HTS, make sure that the voltage on the label which is on the left bottom of the back panel matches the local voltage. Warning Never operate the instrument on an ungrounded electrical outlet.
Page 11
Users can select any file in the table, for example by clicking the "DNA" button to enter the list of programs. Click the program you want to run, and click "Select" to enter the program details interface. The program detail interface displays the step number, board position number, item name, time, volume, temperature and other parameters.
Page 12
Click "Cancel", a dialog box will indicate whether to exit the current running program or not. Click "Cancel" to continue running, and click "Confirm" to stop running. Customers should use this function with caution. 4.2.3 New Program Creation Click "New Program" on the main interface to create a new programs according to the customer’s requirements.
Page 13
Note Purifier HTS only contains 4 plate positions, of which 1 and 4 are heating plate positions. Generally, No. 1 The plate position corresponds to the combination step, and the plate position 4 corresponds to the elution step. Please set the program steps according to the plate position.
Page 14
4.2.4 User Settings The user can select "User Setting" on the main screen to set the parameters of Purifier HTS. After entering the user setting interface, the user can set the language, fan switch, light switch and prompt tone switch.
(1) Press the power switch on the rear panel of the instrument to the OFF position to turn off Purifier HTS. It is recommended to turn off the instrument at night and on weekends. (2) Wipe the surface of the instrument with a soft cloth or paper towel with distilled water, mild detergent (SDS, sodium lauryl sulfate) or soap solution.
Although Purifier HTS is made of high-quality materials, you must remove the spilled salt solution, chemical solvent, acid or alkaline solution from the surface immediately to prevent damage to the instrument.
5.9 Disposal of the instrument If Purifier HTS must be discarded, please follow the specific regulations of the laboratory and the country for the treatment of biological waste. The disposal of the instrument is carried out in accordance with the laws and regulations of the local authorities on the recycle of electronic equipment and waste.
Chapter 6 Technical Parameters 6.1 General Parameters As part of our continuous product development plan, the manufacturer reserves the right to change any parameters without prior notice. General parameter Size Device (510mm x 430mm x 470 mm)±10mm Weiget Device about 27kg[±0.5kg] Operating temperature range + 5°C to + 40°C The maximum relative humidity is 80% when the...
The maximum relative humidity is 80% when the temperature is Humidity lower than 31℃,and the humidity drops linearly to 50% at 40℃ AC power fluctuation Not more than ±10% of the marked voltage 6.4 Compliance Purifier HTS meets the following compliance: YY 0505-2005/IEC 60601-1-2:2001 GB 4793.1-2007/IEC 61010-2:2001 Page 17...
6.5 Error code P96 error code P96 error type Solution If the error is reported X axis photoelectric switch error repeatedly after restarting, contact the engineer. If the error is reported X axis motor error repeatedly after restarting, contact the engineer. If the error is reported Z axis photoelectric switch error repeatedly after restarting, contact...
This instruction is for end-users, such as scientific researchers and laboratory technicians; it provides basic information about the 96-channel automatic nucleic acid extraction and purification instrument (referred to as Purifier HTS), including the installation and operation of the instrument. Page 19...
(4) This equipment can only be operated with specially designed software. (5) Do not force the microplate onto the turntable. (6) Purifier HTS is for laboratory research use only. Please comply with appropriate laboratory safety precautions, such as wearing protective clothing and following certified laboratory safety procedures.
Chapter 8 Manufacturer and Medical Device Information Manufacturer: GENFINE BIOTECH (CHANGZHOU) CO., LTD Medical device production record number: NMPA Reg. No.: 20220032 Manufacturer Address: Building E4, Changyang Road, West Taihu Technology Industrial Park, Changzhou City, Jiangsu Province Contact: 0519-83761557 Page 21...
Purifier HTS(Chinese version) P96S-B01-02 Purifier HTS(English version) P96S-B01-03 Purifier HTS(Custom version) 9.2 List of accessories and consumables Table 9-2 List of accessories and consumables used with Purifier HTS Content Purifier 96 magnetic tip comb 96 channel 96-well plate 96 channel...
Appendix A: Certificate of Decontamination _________________________________________________ Name: Address:_______________________________________________________________ _________________________________________________ Telephone / Fax: Device:______________________________ Record Number:___________________________ A) I confirm that the returned items are not contaminated with liquid, toxic, carcinogenic or radioactive substances or any other harmful substances. B) I confirm that the returned items have been decontaminated and can be processed without affecting the health of the personnel.