  • Page 1 Instructions For Electronic Pipette Ipipette GENFINE BIOTECH (CHANGZHOU)CO., LTD http://www.genfine.com/ 18501955112 marketing@genfine.com...
  • Page 2 Instructions For Electronic Pipette Ipipette A- Pipetting key/ G- Label window Confirmation key H- Pipette shaft B- Adjustment wheel I- Tip cone C- Tip ejection key J- LCD display D- Function key K- Finger rest E- Speed key L- External lithium battery F- Handgrip Ⅰ...
  • Page 3 Instructions For Electronic Pipette Ipipette USB charg jack N - Power key 1000 Battery charg e indicator 3 - Navigation bar - Volume settin 4 - Speed indicator Ⅱ...
  • Page 4 Instructions For Electronic Pipette Ipipette Suction and Discharge Key/Confirm Key/Wake Up Key ---In the main interface, it isresponsible forsuctionand discharge. --When it isnecessary toconfirm, it is the confirm key. --In standbymode, you can wake up the systemto start working. Adjusting Knob ---The volume increases with clockwise rotation and decreases with counterclockwise rotation.
  • Page 5: Table Of Contents

    Instructions For Electronic Pipette Ipipette Contents - Product Over view - Instructions B efore Use Operating Ins tructions - Other Operat ing Instructions - Fault Analysis and Elimination - Maintenance - Attention - Warranty Info rmation...
  • Page 6: Product Over View

    Instructions For Electronic Pipette Ipipette 、 Product Over view 。 ---This product is a brand-new product developed by Genfine Biotechn ology Co.Ltd Move the pipette and operate according to the principle of air displacement. I is safe and accura te It is an ideal tool to transfer liquid efficiently.
  • Page 7: Instructions Before Use

    Instructions For Electronic Pipette Ipipette 2. Instructions Before Use 2.1 Battery Matters Needing Attention: ① It is strongly recommended that the battery be fully charged before using this product. ② Please use the original charger and battery combination. If used other batteries may cause damage to the pipette.
  • Page 8 Instructions For Electronic Pipette Ipipette Install The Battery...
  • Page 9: Operating Instructions

    Instructions For Electronic Pipette Ipipette 2.2 Installation and Push Out of Suction Head Installation of suction head: Before installation, please select the matching suction head. During the installation, the pipette should be inserted into the suction head vertically, rotated left and right for half a circle, and then tightened.
  • Page 10 Instructions For Electronic Pipette Ipipette Turn the knob to adjust the liquid volume Turn the knob counterclockwise Press the suction key to suck Press the suction key Then press the discharge key to return to the automatic suction interface Press the drain key Note: 1.
  • Page 11 Instructions For Electronic Pipette Ipipette 3.2.2 Quick dispensing (STEPPER) In this mode, the user can control the discharge volume and the number of discharges. This pipetting combination is based on the automatic aspiration mode and is discharged in batches during discharging. If you need to dispense 10 times, drain 10μl each time, you...
  • Page 12 Instructions For Electronic Pipette Ipipette Note: Since the quick dispensing mode is based on the extension of the automatic aspiration modeFunction, so when the user is using the quick dispensing process, please reserve a supplement liquid. For example: dispense 80μl, 4 times, please adjust the automatic aspiration volume to90μl, reserve 10μl as...
  • Page 13 Instructions For Electronic Pipette Ipipette Press the aspiration button until the target volume Press and hold the aspiration button Long press the empty button for 2 seconds Press the drain button to drain the liquid Empty liquid and return to manual suction interface 3.2.4 Multiple aspiration (ASPIRATE)
  • Page 14 Instructions For Electronic Pipette Ipipette Press the aspiration button until the target volume Press the aspiration button twice to aspirate 50μl each time, a total of 100μl Press the switch button to change from suction to discharge Press the switch key...
  • Page 15: Other Operating Instructions

    Instructions For Electronic Pipette Ipipette Press the aspiration button to aspirate the total volume Press the aspiration button to aspirate the total volume Finally, press the drain button to drain the remaining liquid from the tip and return to the multiple dispensing interface.
  • Page 16 Instructions For Electronic Pipette Ipipette 3. Unequal dispense The liquid is sucked and discharged several times at a time. For example, 10μl of the first drainage, 20μl of the second drainage, 30μl of the third drainage, 40μl of the fourth drainage, 110μl of liquid can be aspirated and discharged...
  • Page 17: Fault Analysis And Elimination

    Instructions For Electronic Pipette Ipipette 4 Fault analysis and elimination 4.1 Troubleshooting Solution Phenomenon Cause ○ The piston is too tightly -Reinstall the piston ○ The piston is not -Grease the piston lubrication ○ Solidification of the -Remove old grease...
  • Page 18: Maintenance

    Instructions For Electronic Pipette Ipipette Phenomenon Cause Solution ○ Improper calibration -Recalibrate with the Inaccurate ○ Pipette liquids that are problem liquid dispensing easily volatile or density -Pre-wash tips or of special is significantly different recalibrate liquids from water -Use high-quality tips ○...
  • Page 19 Instructions For Electronic Pipette Ipipette ③Disinfection The lower end of the electric pipette adopts corrosion-resistant and h-temperature resist ant materials, which can be sterilized under h temperature and h igh pressure. After cleaning the dismantled ℃, 1bar for 20 minutes. After...
  • Page 20 Instructions For Electronic Pipette Ipipette Cleaning and maintenance Handling Prevention methods Liquid method Disassemble the lower end of the Aqueous pipette, rinse the contaminated solution parts with double distilled water, dry it below 60°C or let it dry buffer naturally, and apply a small amount of grease to the piston.
  • Page 21: Attention

    For warranty details, please contact Genfine Company or your local dealer. Product instructions are subject to update without notice.