TABLE OF CONTENTS Introduction ................... 3 Safety instructions ................4 Product information ................5 OptiCell 2 PRO V2 air mattress system ........... 5 Preparation .................... 5 The OptiTex™ cover ................6 Applications ................... 6 Fire protection ..................6 System features ..................
Thank you for choosing the OptiCell 2 PRO V2 alternating pressure air mattress from Järven Health Care. Our goal is for you to be satisfied with the mattress in terms of both function and service life. Therefore, please take a few minutes to read through this guide to understand the best way to manage and care for your mattress.
SAFETY INSTRUCTIONS Read the following information before using your OptiCell mattress! • The product should only be used in accordance with this user guide. • The product should not be assembled, combined or repaired with any parts or accessories other than those described in this guide or in other documentation from Järven Health Care.
PRODUCT INFORMATION OPTICELL 2 PRO V2 AIR MATTRESS SYSTEM The product is designed to help prevent and treat pressure ulcers from Categories 1-3 (low risk to very high risk) for patients within the weight range 0-230 kg. The product is intended for use with a sheet between the patient’s skin and the mattress surface.
It aids the prevention and treatment of pressure ulcers up to and including Category 3. OptiCell 2 PRO V2 has a height of 18cm and comes with a durable OptiTex cover and a CPR valve for emergency situations.
SYSTEM FEATURES The OptiCell 2 PRO V2 air mattress is constructed so that individual cells can be disconnected, even while a patient is lying on the mattress. The mattress has three different types of cells: • Head cells – 13cm high continuously inflated air cells that are positioned under the patient’s head.
STORAGE The OptiCell 2 PRO air mattress system should be stored in the dry and at room temperature. We recommend that the mattress be kept rolled up in one of our optional transport bags.
The mattress can be used immediately after delivery and unpacking. Inflating the mattress takes about 15 minutes and should be done before the patient is placed on the mattress. AUTOMATIC ADJUSTMENT AND ALTERNATING PRESSURE The pump will automatically adjust the air pressure based on the patient’ s weight, length, location, movements and position on the mattress.
AIR HOSE CONNECTION Connecting the air hoses to the pump: To remove the protective caps and to connect or disconnect the air hose, push in the two contact clips on the side of the connector. This action releases the hooks that hold the connector in place.
HAND CHECK TEST Test the mattress’s function with a simple hand check. To ensure that the mattress is providing the right therapeutic function, a simple hand check is performed. The test should be performed each time a mattress is taken into operation in order to detect any problems with the mattress’s air pressure.
INSPECTION AND TROUBLESHOOTING INSPECTION OF YOUR OPTICELL AIR MATTRESS SYSTEM An inspection plan should be established in order for your air mattress to function and last for as long as possible. Please contact us if you would like assistance with this. Järven Health Care’s contact details can be found on the back cover of this guide.
CLEANING CLEANING THE PUMP To clean the pump, use a damp cloth and mild detergent on the pump housing and the power cable. CLEANING THE OPTITEX COVER The cover is made of a polyurethane (PU) material on both top and bottom. The entire cover (both upper and lower parts) can be cleaned with disinfectants.
PUMP OVERVIEW – OC20 V2 The power switch is located on the right side of the pump. Press “On/Off” to start or stop the pump. READY Inflating the mattress takes approximately 15 minutes. When the “Ready” lamp is lit, the mattress is ready to use. COMFORT There are six comfort modes to choose from.
AIR PRESSURE STATUS This indicator shows when the air pressure is normal or too low. When the air pressure is correct, the “Normal” lamp will be lit. If the “Low” lamp lights up, the air pressure is too low and the mattress should not be used. STATIC/DYNAMIC Press this button to select “Static”...
RECYCLING All materials in your OptiCell air mattress system, with the exception of certain parts of the pump, are energy-recoverable by incineration at designated combustion plants. PUMP An OptiCell pump must not be disassembled, but should be disposed at a recycling station and sorted as ”Electronic waste”.
OTHER INFORMATION PRODUCT LIABILITY Product liability legislation (EU Directive 85/374/EEC, Swedish Law SFS 1992: 18) establishes the manufacturer’s, importer’s or seller’s statutory liability for any damage caused by the product. The regulations apply to physical products in the form of movables, even if incorporated into another movable or into an immovable.
PRESSURE ULCER PREVENTION WORK A clinical assessment of the patient’s risk of developing pressure ulcers should be carried out. Aids to the assessment include the Norton scale, Braden scale and Risk Assessment Pressure Sore (RAPS) scale. The assessment must be done on arrival or at the beginning of a care period, regardless of whether the patient is in a care facility or cared for in his or her own home.