INDICATIONS
Philips air hoses are intended for use by, or
under the supervision of, a licensed physician
or other healthcare provider for the non-
invasive measurement of neonatal, pediatric,
and adult human blood pressure.
SYMBOL DEFINITIONS
The following are definitions for symbols that
might appear on package labels for the
products described in this document:
Quantity per package
Follow operating instructions
Not manufactured with natural
rubber latex
Not manufactured with
polyvinyl chloride (PVC)
Not manufactured with
Di(2-ethylhexyl)phthalate
(DEHP)
Caution
Storage temperature range
Storage humidity range
Prescription use only
ENGLISH
WARNINGS
• Use neonatal and infant #5 cuffs only with
hoses designated for Neonatal use (see
cover and Select Appropriate Hose).
• Use pediatric and adult cuffs only with
hoses designated for Adult use (see cover
and Select Appropriate Hose).
• Hoses are not designed to be sterilized. Do
not autoclave.
• Prevent cleaning/lubricating fluids from
entering hose. Fluids can damage monitor
equipment.
• Lubricate the cuff-end connector of
M1598B, M1599B, Mobile CL, and Efficia
hoses with a non-silicone spray release
agent when it becomes difficult to connect
to cuffs. Do not lubricate the instrument-
end connectors. Do not lubricate the
connectors for the M1596C and M1597C
hoses.
• Philips NIBP air hoses use connectors
which have the potential to misconnect
with small-bore connectors used in non-
NIBP healthcare applications. When using
the Philips NIBP air hoses, ensure that
connectors are properly connected to
Philips NIBP devices only.