BVI 9400 Calibration Quick Reference
BVI 9400 Calibration Quick Reference
The entire calibration process may take up to 15 minutes. You may calibrate your instrument as often as you like; you do not have
to wait for the next scheduled calibration date. However, at a minimum, you must calibrate the instrument every 12 months.
www.verathon.com https://my.scanpoint.com
For more information, call 800.331.2313 (in the U.S. and Canada) or contact your local Verathon Medical
Verathon Medical

Prepare for Calibration

Set up the calibration tank in a location that is within
10 feet of Charger and Wireless Hub.
Place the calibration tank on a flat, nonreflective surface and
remove the lid. Pour clean, room temperature water into the
container, filling to the indicator mark. Make sure the water
contains a minimal amount of bubbles. Place the spiral-
shaped target in the container, using the notches to position
target correctly. Replace the lid on the calibration container.

Place the probe

Place the Probe in the cutout in the top of the calibration
tank. Make sure the black tip of the Probe is submerged
in the water.

Open ScanPoint

QuickPrint
®
From your desktop, open QuickPrint by double-clicking
the QuickPrint icon.
Establish BVI 9400 - QuickPrint
connection
Press the ScanPoint
button on the BVI 9400.
®
On the QuickPrint screen, click the FIND NEW button.
When QuickPrint establishes a connection with the
BVI 9400, an icon will appear along the left menu.
®
are registered trademarks of Verathon Inc. in the US and other countries. Copyright © 2007 by Verathon Inc. All rights reserved.
ScanPoint
®
button
®
representative. ScanPoint
®
, BladderScan
®
, Verathon
®
, and

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BVI 9400 and is the answer not in the manual?

Questions and answers

Summary of Contents for BLADDERSCAN BVI 9400

  • Page 1 On the QuickPrint screen, click the FIND NEW button. When QuickPrint establishes a connection with the BVI 9400, an icon will appear along the left menu. www.verathon.com https://my.scanpoint.com For more information, call 800.331.2313 (in the U.S. and Canada) or contact your local Verathon Medical ®...
  • Page 2 Probe’s LCD screen. Do not disconnect the Charger and Wireless Hub from the computer or otherwise break the connection between the BVI 9400 and the ScanPoint host computer ® until all processing is complete.

Table of Contents