Advertisement

HEARING AID
Instruction Manual Model: JH-W5
Please read this user manual thoroughly and
Life cycle of product is three years
completely before using the product.
VER: W5(A620)-20211213-V01

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the JH-W5 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Jinghao JH-W5

  • Page 1 HEARING AID Instruction Manual Model: JH-W5 Please read this user manual thoroughly and Life cycle of product is three years completely before using the product. VER: W5(A620)-20211213-V01...
  • Page 2: Table Of Contents

    7.HOW TO USE THE CHARGING CASE CONTENTS V.MAINTENANCE AND CLEANING I.IMPORTANT SAFETY INSTRUCTIONS VI.HEARING AID CARE 1.CONTRAINDICATION VII.TROUBLESHOOTING GUIDE 2.WARNINGS VIII.TIPS FOR BETTER COMMUNICATION 3.NOTICE IX.TECHNICAL SPECIFICATION II.DO NOT USE IF ANY OF THE FOLLOWING APPLY TO YOU X. ACCESSORIES LIST III.MATTERS NEEDING ATTENTION, WARNINGS AND TIPS XI.NORMALIZED SYMBOLS IV.HOW TO USE YOUR DEVICE...
  • Page 3: I.important Safety Instructions

    - Routine maintenance to keep the device clean is required. I.IMPORTANT SAFETY INSTRUCTIONS Read all the information in this instruction manual and any Ensure that ear domes are free of debris and moisture at all other literature included in the box before using the device. times.
  • Page 4 magnetic fields. Specific warnings / statements - Do not attempt to open or repair the devices yourself. Damage can result and will void the warranty. Keep small parts away from and out of children´s reach to - Do not store your devices in a hot/humid locations. You avoid accidental ingestion should never leave your hearing aids near heating appliances Warning not to use in explosive or oxygen-enriched...
  • Page 5: Ii.do Not Use If Any Of The Following Apply To You

    III. MATTERS NEEDING ATTENTION, WARNINGS AND TIPS II. DO NOT USE IF ANY OF THE FOLLOWING APPLY TO YOU: 1. Do not immerse the hearing aid in any liquid. Remove the 1) Congenital atresia of external ear or deformity of any part of hearing aid before swimming or bathing.
  • Page 6: Iv.how To Use Your Device

    Step 2: Push the ear dome rim membrane outward. IV.HOW TO USE YOUR DEVICE Step 3: Hold on to the end of the hearing aid and push firmly to ensure that the dome is fastened securely. 1. Volume Control 2. Microphone 3.
  • Page 7: Left & Right Hearing Aid Markings

    Step 2: Move the body slightly into auricle for holding. 2.【LEFT & RIGHT HEARING AID MARKINGS】 There is a L or R marking on the hearing aid. This will tell you if it is a left or a right hearing aid. 3.2 Removing the hearing aid L marking for left hearing aid.
  • Page 8: Turn On/Turn Off

    *Note: 4.【TURN ON/TURN OFF】 Turn On: Take the hearing aid out of the charging case, the 1. It may be in few circumstances that the ear dome remain hearing aid will enter the standby status. The hearing aid will attached to the auditory meatus after the hearing aid is taken turn on automatically in 3 seconds when you wear the hearing out.
  • Page 9: Volume Adjustment

    Note: The hearing aid built in with ‘delay turn on’ function, after placing correctly in the ear canal, automatically turn on after 3 seconds delay from the ‘beep’ tone. 5.【VOLUME ADJUSTMENT】 To increase the volume, turn the volume control towards the "+"...
  • Page 10: Hearing Aid Charging

    6.【HEARING AID CHARGING】 Put the hearing aid back into the correct slot of the charging case, they will automatically turn off and start charging. Orange light will indicate hearing aids are charging. After fully charged, they will light in green and the light turn off in dicates charging is complete.
  • Page 11 7.5 Charging case Turn ON/OFF 7.4 The charging process will automatically stop when the case Press and hold the lower part of button for 5 seconds until the and hearing aids are fully charged, so the hearing aids can be solid orange light indicates the charging case is off.
  • Page 12: V.maintenance And Cleaning

    4.When the hearing aid and case is not functioning properly, To avoid potential damage from excess humidity or please check the “troubleshooting guide”. If none of the perspiration, the use of a drying kit is also recommended. solutions work, please contact our after-sales service team. 2.
  • Page 13 or shelf rather than the bathroom or kitchen. VI. HEARING AIDS CARE - To ensure the longest life of your rechargeable hearing aids - Keep your hearing aids clean. Extreme temperatures (heat or batteries and the batteries in the charger: cold), moisture, and foreign substances can result in poor - Fully charge hearing aids batteries every night.
  • Page 14: Vii.troubleshooting Guide

    VII.TROUBLESHOOTING GUIDE Hearing has changed Contact your physician to report any sudden drop in hearing SYMPTOM POSSIBLE CAUSES SOLUTIONS Hearing Aid plays Indication for low battery Charge hearing aid Incorrect wearing, the in Remove the hearing aid, wear notification sound ear detection sensor failure it correctly Volume too low...
  • Page 15: Viii.tips For Better Communication

    VIII.TIPS FOR BETTER COMMUNICATION [Fault self-check] Possible Causes Solution project It will take practice, time, and patience for your brain to adapt Poor contact between Reinsert the hearing aid into the to the new sounds that your hearing aids provide. Hearing is hearing aid and charging case charging case...
  • Page 16: Ix.technical Specification

    IX.TECHNICAL SPECIFICATION - Develop realistic expectations of what your hearing aids OSPL Max. ≤120dB+3dB can and cannot do FOG Gain ≤35dB+3dB - Better hearing with hearing aids is a learned skill combin- EQ Input Noise ≤29dB+3dB ing desire, practice, and patience. You will learn and adjust, it 450Hz~4000Hz Frequency Range just takes a little time.
  • Page 17: Xi.normalized Symbols

    XI. NORMALIZED SYMBOLS Notes:This hearing aid is produced under the strict quality system of Huizhou JINGHAO Medical Technology Co.,Ltd. Follow operating instructions The device fulfils the provisions of the EC directive 93/42/EEC B type applied part (Medical Device Directive) and the Harmonize Standard as IEC...
  • Page 18: Xiii.warranty

    All other damage claims excluded. A warranty claim must be submitted with the purchase receipt. Huizhou Jinghao Medical Techology Co., Ltd. Address: Floor 6, Huicheng Industry Building, No.9 Huifeng Dong’er Road, Zhongkai High-tech Zone, Huizhou City, Guangdong Province, China.