Quick Reference Guide
Quick Reference Guide
Quick Reference Guide
Quick Reference Guide

SPECIFICATIONS

• RANGE

EC
0.001 to 9.999 µS/cm*, 10.00 to 99.99 µS/cm,
100 to 999.9 µS/cm, 1.000 to 9.999 mS/cm,
10.00 to 99.99 mS/cm, 10.00 to 99.99 mS/cm,
100.0 to 1000.0 mS/cm
Actual conductivity 1000.0 mS/cm
Shows values up to 1000.0 mS/cm
Resistivity
1.0 to 99.9 ohms, 100 to 999 ohms,
1.00 to 99.9 Kohms, 10.0 99.9 Kohms,
100 to 999 Kohms, 1.00 to 9.99 Mohms,
10.0 to 100.0 Mohms*
TDS
0.00 to 999.9 ppm
1.000 to 400.0 g/L
Salinity
% NaCl
0.0 to 400.0 % NaCl
Sea Water Scale
0.00 to 80.00 ppt
Practical Salinity
0.01 to 42.00 (PSU)
Temperature
–20.0 to 120.0 °C (-4.0 to 248.0 °F)

• ACCURACY

EC
±1% of reading (±0.01 µS/cm or 1 digit)
Resistivity
±1% of rdg (10 ohm or 1 digit)
TDS
±1% of reading (±0.05 ppm or 1 digit)
Salinity
±1% of reading
Temperature
±0.2 °C

OPERATIONAL GUIDE

• To switch ON and OFF: press ON/OFF ( ).
• To take measurements: with meter off, connect the electrode's DIN connector
to appropriate socket, remove the protective cap and immerse the electrode in the
solution with the sleeves holes completely submerged. Switch on the meter. Stir
briefly and wait for a stable reading. For accurate readings tap the probe
repeatedly to remove any air bubbles that may be trapped inside the sleeve.
• To select the range, conductivity, resistivity TDS or salinity: press RANGE.
* The 0.001 µS/cm EC range and 100 MOhms Resistivity range are not available with the 4 m cable probe.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 98192 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments 98192

  • Page 1 HI 98192 USP Compliant EC/TDS/NaCl/Resistivity/Temperature Portable Meter Quick Reference Guide Quick Reference Guide Quick Reference Guide Quick Reference Guide Quick Reference Guide SPECIFICATIONS • RANGE 0.001 to 9.999 µS/cm*, 10.00 to 99.99 µS/cm, 100 to 999.9 µS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 10.00 to 99.99 mS/cm,...