Page 1
IRT225 10:1 Mini IR Thermometer Introduction Congratulations on your purchase of the IRT225 10:1 infrared thermometer The IRT225 can quickly and accurately determine the surface temperature by measuring the infrared energy radiated from the target surface. It is suitable for non-contact surface temperature...
Page 2
Safety Warning: LASER RADIATION DO NOT STARE INTO BEAM OUTPUT < 1mW COMPLIES WITH EN60825-1:2014 To avoid eyes or personal injury, please read the following safety instructions before using the thermometer: Please do not irradiate people or animals with laser directly or indirectly.
Page 3
Description METER DESCRIPTION Laser Infrared sensor Trigger Battery cover °C/°F switching button Battery compartment LCD DESCRIPTION Function Indicator...
Operation Viewing the Last Measured Value In the off state, short press the trigger to turn on the thermometer, it will display the last measurement data that was held before the last shutdown. Temperature Measurement Aim the thermometer at the target, then pull and hold the trigger, the laser indicator on the LCD display shows that the laser is turned on and points to the target, the measured temperature value will be displayed on the LCD, and the maximum or minimum temperature of the target area can be...
Field of View Make sure that the measured target is larger than the spot diameter. The smaller the target is, the closer the test distance should be (please refer to D:S for the spot diameter at different distances). To obtain the best measurement result, it is recommended that the target being measured is 2.5 times larger than the spot diameter.
Troubleshooting Symptom Problem Action appears Measured value when greater than Stop measuring measuring maximum range -OL appears Measured value is less when Stop measuring than the minimum range measuring Place thermometer in a Exceeding the minimum 0°C~50°C appears or maximum operating (32°F~122°F) when booting ambient temperature...
Low battery, emissivity mismatching, measured Please refer to the distance far, instructions measurement measured target field of view, D:S, is inaccurate diameter is less than etc. 20mm, etc. Maintenance Cleaning and storage 1. The meter should be wiped with a damp cloth and mild detergent when necessary.
Specifications Temperature -20°C~400°C (-4°F~752°F) range >0°C: ±2.0°C or ±2% of reading, whichever is Accuracy greater Ambient ≤0°C: ±(2.0°C+0.1°C/°C) temperature >32°F: ±4.0°F or ±2% of reading, whichever is 21°C~25°C greater (70°F~77°F) ≤32°F: ±(4.0°F+0.1°F/°F) Temperature ±0.1°C/°C or ±0.1%/°C, whichever is greater coefficient (±0.1°F/°F or ±0.1%/°F, whichever is greater) D:S ratio 10:1 (calculated at 90% energy)
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase.
Need help?
Do you have a question about the IRT225 and is the answer not in the manual?
Questions and answers