Single-Use PDF Data Logger
Appearance
6
5
2
1
4
3
7
1. USB Protective Cover
4. LED Indicator
2. LCD Screen
5. Light Sensor
3. Button
(1)
6. Humidity Sensor
7. Shelf life
Before you begin
·ElitechLog Software download link: www.elitechlog.com/softwares
·Online configuration site link:
Technical Specifications
Recording Options
Single Use
-30°C ~ 70°C, 0%RH ~ 100%RH
Temperature Range
±0.5°C(-20°C ~ +40°C), others ±1.0°C
Temperature Accuracy
 
±0.3°C(-30℃ ~ +70°C) - for LogEt 1Bio only
Humidity Accuracy
±3%RH(20%RH ~ 80%RH), others ±5%RH -
for LogEt 1TH only under 25°C
Resolution
0.1°C , 0.1%RH
Max. 16,000 points
Data Storage Capacity
2 years/CR2450 button cell
(3)
Shelf Life / Battery
Recording Interval
12 minutes (Default, others on request)
Recording Duration
Up to 120 days (Default, others on request)
Start Mode
Button or software
Stop Mode
Button, software or stop when full
IP67 (Not for LogEt 1TH)
Protection Class
Via Elitech software
Reprogrammable
Via Online Configuration Web
01
www.elitechlog.com
Note:
Button
(1)
Function Instructions:
Operation
Function
Status Indication
Start Recording:
Start/Stop
Press & Hold
Recording
5 Seconds
Stop Recording:
Light Up LCD
See Indication
Backlight;
Single Click
Instructions
Page Up/Down
Mark Success:
Double Click
Mark Events
Mark Failure:
(2)
The code like
will shows on logger's LCD screen to indicates
status. The red square
indicates the logger's red LED light is flashing;
the green square
indicates the green LED light is flashing. Each
individual square indicates how many times it will blinks and two
adjoining squares like indicates both red and green lights blink
simultaneously. The same rules applied below.
EN12830, CE, RoHS
Certifications
As Hardcopy
Validation Certificate
ElitechLog Data Management Software
Software
Win(V4.0.0 or newer)
/ElitechLog Mac (V1.0.0 or newer)
Mac OS 10.10 or higher
Compatible OS
Windows XP/7/10
Automatic PDF report
Report Generation
Software Password Protection
Password Protection
USB 2.0 (Standard Type A Connector)
Connection Interface
(4)
Optional, Up to 5 thresholds
Alarm Configuration
Note:
(3)
Depending on optimal storage conditions (15°C to 23°C / 45% to 75% RH)
(4)
Depending on application temperature (very low/high temperatures may
shorten it)
Software download: www.elitechlog.com/softwares
Indication Instructions
1.LCD Screen Indication
(2)
1
2
3
1. Alarm status
2. Working status
3. Humidity/Logging Intervals/Recorded Points
4.Loop mark
Note:
(5)
Only displayed when Alarm is enabled.
(6)
Indicates current alarm status, e.g. if temperature goes above AH1 setting, AH1 code will appears on LCD screen.
2. Other LCD Page Indications:
Single click the Button can browse each LCD page.
1. Current Temperature & Humidity
4. Minimum Temperature & Humidity
3. Meaning of LED blinks
Single click the Button make LED light blinks you may check logger status based on it.
LED blinks as...
www.elitechlog.com
8
7
6
5
4
(5)
5. Temperature display
6. Function indication
7. Battery indicator
8.Alarm threshold
(6)
2. Recorded Points
3. Maximum Temperature & Humidity
5. System Date: Month-Day
6. System Time: Hour:Minute
Indicates status...
Not Started
Delayed Start/Timing Start
Started - OK
Started - Alarm
Stopped - OK
Stopped - Alarm
Software download: www.elitechlog.com/softwares
02

Advertisement

loading

Summary of Contents for Elitech LogEt 1

  • Page 1 User Manual LogEt 1/LogEt 1TH/LogEt 1Bio Single-Use PDF Data Logger Appearance Indication Instructions Note: Button Function Instructions: 1.LCD Screen Indication Operation Function Status Indication Start Recording: Start/Stop Press & Hold Recording 5 Seconds Stop Recording: Light Up LCD See Indication Backlight;...
  • Page 2 If battery indicator icon on LCD screen only display half of the icon, please do not use the logger for long-distance transport. LogEt 1 series should avoid contact with unstable chemical solvents or other organic compounds especially should avoid for long-term storage or exposed to environments that contain high concentrations of ketene, acetone, ethanol, isopropanol, toluene and etc.

This manual is also suitable for:

Loget 1thLoget 1bio