Cleaning
The front panel membrane conceals and protects switches and LEDs. It may be wiped clean using a
damp cloth. Be sure to disconnect the unit from the mains before cleaning, and do not use any
abrasive cleaners or organic solvents.
Periodic testing
Check the mains lead and plugs at both ends for any damage.
Periodically check the earth bonding and insulation, by performing a Portable Appliance Test (PAT).
Most schools and local authorities have a regular schedule for such testing.
Check the action of the electronic cutout, by short circuiting the power supply output (check dc and ac
outputs separately) using a short 4mm plug lead.
Check that the fuse in the mains plug (5A recommended) and the two fuses (active and spare) in the
IEC socket on the back panel are all of the correct rating.
Disposal of Waste Electrical and Electronic Equipment (WEEE)
Warranty, repairs and spare parts
The power supply is guaranteed for a period of one year from the date of delivery to the customer.
This warranty does not apply to defects resulting from the action of a user such as misuse, improper
wiring, any operations outside of its specification, improper maintenance or repair, or unauthorized
modification.
Our liability is limited to repair or replacement of the product. Any failure during the warranty
period should be referred to Customer Services.
In the event of a fault, apart from replacing the instrument fuse in the IEC socket, the power supply
should be referred to the Philip Harris recommended repair agent.
Please contact Customer Services or
Do not dispose of this product with household waste
For the proper treatment, recovery and recycling please take
this product to an appropriate collection point.
If you are unsure where this is, contact your Local Authority
By disposing of this product correctly you will be providing
positive help to the environment.
techsupport@philipharris.co.uk
for advice
Need help?
Do you have a question about the Retro Basic and is the answer not in the manual?